65W Single Output Medical Type
MPS-65 series
Features :
Universal AC input / Full range
Low leakage current 0.3mA
Protections: Short circuit / Overload / Over voltage
Cooling by free air convection
100% full load burn-in test
Fixed switching frequency at 45KHz
3 years warranty
SPECIFICATION
MODEL
MPS-65-3.3 MPS-65-5
MPS-65-7.5 MPS-65-12 MPS-65-13.5 MPS-65-15 MPS-65-24 MPS-65-27 MPS-65-48
DC VOLTAGE
3.3V
5V
7.5V
8A
12V
13.5V
15V
24V
27V
48V
RATED CURRENT
CURRENT RANGE
RATED POWER
12A
12A
5.2A
4.7A
4.2A
2.7A
2.4A
1.35A
0 ~ 1.5A
64.8W
0 ~ 15.2A
39.6W
0 ~ 13.8A
60W
0 ~ 9.6A
60W
0 ~ 6A
62.4W
0 ~ 5.4A
63.45W
0 ~ 4.8A
63W
0 ~ 3A
64.8W
0 ~ 2.7A
64.8W
OUTPUT POWER (max.)
72W(+3.3V:50W;+5V:69W)with 18CFM min. Forced air convection
OUTPUT
RIPPLE & NOISE (max.) Note.2 80mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
VOLTAGE ADJ. RANGE
2.97 ~ 3.63V 4.5 ~ 5.5V
6.75 ~ 8.25V 10.8 ~ 13.2V 12.2 ~ 14.85V 13.5 ~ 16.5V 21.6 ~ 26.4V 24.3 ~ 29.7V 43.2 ~ 52.8V
VOLTAGE TOLERANCE Note.3 3.0%
3.0%
1.0%
3.0%
3.0%
1.0%
3.0%
2.0%
1.0%
2.0%
2.0%
1.0%
2.0%
2.0%
1.0%
2.0%
2.0%
1.0%
2.0%
2.0%
1.0%
2.0%
2.0%
1.0%
2.0%
LINE REGULATION
LOAD REGULATION
SETUP, RISE TIME
HOLD UP TIME (Typ.)
1.0%
3.0%
800ms, 30ms/230VAC
800ms, 30ms/115VAC at full load
16ms/115VAC at full load
127 ~ 370VDC
50ms/230VAC
90 ~ 264VAC
47 ~ 440Hz
66%
VOLTAGE RANGE
FREQUENCY RANGE
EFFICIENCY(Typ.)
74%
76%
0.9A/230VAC
77%
78%
79%
80%
80%
80%
INPUT
1.6A/115VAC
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
COLD START 17A/115VAC
<0.3mA / 264VAC
35A/230VAC
LEAKAGE CURRENT
73 ~ 105W (3.3V:51 ~ 75W)(5V:70 ~ 105W) rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVERLOAD
PROTECTION
3.8 ~ 4.46V 5.75 ~ 6.75V 8.63 ~ 10.1V 13.8 ~ 16.2V 15.5 ~ 18.2V 17.25 ~ 20.25V 27.6 ~ 32.4V 31 ~ 36.45V 55.2 ~ 64.8V
Protection type : Hiccup mode, recovers automatically after fault condition is removed
-10 ~ +60 (Refer to "Derating Curve")
OVER VOLTAGE
WORKING TEMP.
20 ~ 90% RH non-condensing
WORKING HUMIDITY
-20 ~ +85 , 10 ~ 95% RH
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
0.04%/ (0 ~ 50
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
VIBRATION
ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1 approved
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC
SAFETY STANDARDS
WITHSTAND VOLTAGE
SAFETY &
ISOLATION RESISTANCE
I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25 / 70% RH
EMC
(Note 4)
EMC EMISSION
EMC IMMUNITY
MTBF
Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3
Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2, medical level, criteria A
359.7Khrs min. MIL-HDBK-217F (25
127*76*42mm (L*W*H)
)
OTHERS
NOTE
DIMENSION
PACKING
0.23Kg; 54pcs/14.6Kg/1.35CUFT
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.
(as available on http://www.meanwell.com)
5. Mounting holes M1 and M2 should be grounded for EMI purposes.
6. Heat Sink HS1,HS2 can not be shorted.
File Name:MPS-65-SPEC 2011-12-15