MENT1150
Universal Input 150 Watt Series
Medical / ITE Switch-Mode Power Supply
3
Year Warranty
•100-240VAC Universal Input
•
Meets EISA2007, CEC Efficiency Level V,
EU (EC) No 278/2009 Phase II
•
•
•
•
•
•
•
Desktop Style with LED Power-On indicator
12V to 48V Single Output Models, up to 150W
Modified and Custom Designs Available
Regulated Output with Low Ripple
Impact-Resistant Polycarbonate Enclosure
No Load Power Consumption <0.5W
Meets UL/EN/IEC60601-1 & 60950-1 2nd edition
RoHS
100-240Vac, +/-10%, 47-63 Hz, 1∅
Specifications
All Specifications are typical at nominal input, full load at 25°C unless otherwise stated.
MTBF
> 100K hours (calculated, 25 C at full load)
20 ms min. @ 115VAC
AC Input
Input Current
Inrush Current
Input Fuse
115VAC: 3A, 230VAC: 1.5A Hold-up Time
75A peak @ 264Vac, cold start Overvoltage Protection
130% of nominal output voltage
120-180% of full rated load, auto-recovery
Provided with no damage to supply
Latched, requires input recycling
Meets IEC61000-3-2
4.0A, 250V Internal Primary Current Fuse provided
Overload Protection
Short Circuit Protection
Overtemperature Protection
Input Harmonics
Efficiency
Meets International Efficiency Level V
See chart
Output Voltage
Output Power
Ripple and Noise
Line/Load Regulation
Transient Response
See chart
1% pk-pk max., 20MHz BW
Including Cord: Line: +/- 1%, Load: +/-5%
Safety Standards IEC/EN/UL60601-1 & EN/IEC/UL60950-1 2nd edition
Switching Frequency
Isolation
40kHz to130kHz , variable
Maximum deviation of 5% with 25% to
75% load change and 0.2A/uSec slew rate
Not required
Input-Output: 4000VAC,
Input-GND: 1500VAC, Output-GND: 700VDC
Minimum Load
Operating Temperature
0° to 40°C Full load Derate linearly to
40% of full rated load at 70°C
0.04% per °C, maximum
Case Material
Black 94V0 Polycarbonate
see mechanical drawing
Temperature Coefficient
Storage Temperature
Relative Humidity
Case Dimensions
-40°C to +85°C
Altitude
-500 to 10,000 ft, operating; -500 to 40,000 ft, non-operating
5% to 95%, non-condensing
EMI/EMC Compliance
Conducted Emissions
Radiated Emissions
FCC Part 15 Subpart B, Class B, EN55011 Class B, EN55024
FCC Part 15 Subpart B, Class B, EN55011 Class B, EN55024
EN61000-3-3 Complies (dmax<6%)
Voltage Fluctuations
Static Discharge Immunity
Radiated RF Immunity
EFT/Burst Immunity
EN61000-4-2, 6kV Contact Discharge, 8kV air discharge. Criteria A
EN61000-4-3, 3V/m. Criteria A
EN61000-4-4, 2kV/5kHz. Criteria A
Line Surge Immunity
EN61000-4-5, 1kV differential, 2kV common-mode, Criteria A
EN61000-4-6, 3Vrms, Criteria A
Conducted RF Immunity
Power Frequency Magnetic Field Immunity
Voltage Dip Immunity
EN61000-4-8, 3A/m, Criteria A
EN61000-4-11, 0 % Vnominal, 0.5 cycle; 40% Vnominal, 5 cycles;
70% Vnominal, 25 cycles: Criteria A
Line Frequency Harmonics
EN61000-3-2, Class A, B, C & D
______________________________________________________________________________________________________________________________ 1 of 2
SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: info@slpower.com • www.slpower.com