45W Triple Output Medical Type
MPT-45 series
Features :
Universal AC input / Full range
Low leakage current 0.3mA
Protections: Short circuit / Overload / Over voltage
Cooling by free air convection
100% full load burn-in test
Fixed switching frequency at 45KHz
3 years warranty
SPECIFICATION
MODEL
MPT-45A
MPT-45B
MPT-45C
OUTPUT NUMBER
CH1
CH2
CH3
CH1
5V
CH2
CH3
CH1
5V
CH2
CH3
DC VOLTAGE
5V
12V
-5V
12V
-12V
15V
-15V
RATED CURRENT
CURRENT RANGE
RATED POWER
3A
2A
0.3A
3A
2A
0.3A
3A
1.6A
0.3A
0.4 ~ 5A
40.5W
0.2 ~ 2.5A
0 ~ 0.5A
0.4 ~ 5A
42.6W
0.2 ~ 2.5A
0 ~ 0.5A
0.4 ~ 5A
43.5W
0.2 ~ 2.3A
0 ~ 0.5A
OUTPUT POWER (max.)
52W with 18CFM min. Forced air convection
OUTPUT
RIPPLE & NOISE (max.) Note.2 60mVp-p
120mVp-p
60mVp-p
60mVp-p
120mVp-p
100mVp-p
60mVp-p
120mVp-p
100mVp-p
VOLTAGE ADJ. RANGE
CH1:4.5 ~ 5.5V
VOLTAGE TOLERANCE Note.3 4.0%
7.0%
2.0%
4.0%
5.0%
1.0%
1.0%
4.0%
1.0%
3.0%
7.0%
2.0%
4.0%
5.0%
1.0%
1.0%
4.0%
1.0%
3.0%
7.0%
2.0%
4.0%
5.0%
1.0%
1.0%
LINE REGULATION
LOAD REGULATION
SETUP, RISE TIME
HOLD UP TIME (Typ.)
1.0%
3.0%
800ms, 20ms/230VAC
800ms, 20ms/115VAC at full load
12ms/115VAC at full load
127 ~ 370VDC
80ms/230VAC
90 ~ 264VAC
47 ~ 440Hz
73%
VOLTAGE RANGE
FREQUENCY RANGE
EFFICIENCY(Typ.)
75%
75%
INPUT
1.2A/115VAC
0.7A/230VAC
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
COLD START 20A/115VAC
<0.3mA / 264VAC
40A/230VAC
LEAKAGE CURRENT
53 ~ 75W rated output power
OVERLOAD
Protection type : Hiccup mode, recovers automatically after fault condition is removed
5.75 ~ 6.75VDC on CH1
PROTECTION
OVER VOLTAGE
Protection type : Hiccup mode, recovers automatically after fault condition is removed
-10 ~ +60 (Refer to "Derating Curve")
WORKING TEMP.
20 ~ 90% RH non-condensing
WORKING HUMIDITY
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
-20 ~ +85 , 10 ~ 95% RH
ENVIRONMENT
0.04%/ (0 ~ 50 ) on +5V output
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1 approved
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC 1min.
I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25 / 70% RH
SAFETY STANDARDS
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
EMC EMISSION
SAFETY &
EMC
(Note 4)
Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3
Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2, medical level, criteria A
EMC IMMUNITY
MTBF
271.5Khrs min. MIL-HDBK-217F (25
127*76*28mm (L*W*H)
)
OTHERS
NOTE
DIMENSION
PACKING
0.21Kg; 72pcs/18Kg/1.35CUFT
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.
(as available on http://www.meanwell.com)
5. Mounting holes M1 and M2 should be grounded for EMI purposes.
6. Heat Sink HS1,HS2 can not be shorted.
File Name:MPT-45-SPEC 2011-12-15