5秒后页面跳转
HMI63-S480125 PDF预览

HMI63-S480125

更新时间: 2024-09-16 05:36:19
品牌 Logo 应用领域
HITRON 开关医疗输出元件输入元件
页数 文件大小 规格书
2页 139K
描述
UNIVERSAL INPUT AC-DC MEDICAL APPLICATION OPEN FRAME SINGLE OUTPUT 63 WATTS INTERNAL SWITCHING POWER SUPPLIES

HMI63-S480125 数据手册

 浏览型号HMI63-S480125的Datasheet PDF文件第2页 
H HiTRON  
UNIVERSAL INPUT AC-DC MEDICAL APPLICATION OPEN FRAME  
SINGLE OUTPUT 63 WATTS INTERNAL SWITCHING  
POWER SUPPLIES HMI63 SERIES  
FEATURES:  
„ ACCOMMODATE UNIVERSALAC SOURCES  
„ COMPACT 4”x2”x1”PCB FORMAT  
„ MEET MEDICAL STANDARD EN 60601-1/  
UL 2601-1  
„ EMI MEET EN 55011 & FCC CLASS B  
„ CE MARKING COMPLIANCE  
SPECIFICATION  
OUTPUT SPECIFICATION  
Output Voltage: See Ratings Chart.  
Output Current: See Ratings Chart.  
Output Connector: V-M connector.  
Output Wattage: Typ. 62 watts.  
INPUT SPECIFICATION  
Input Voltage: Range typ. 90-264Vac.  
Rating 100 to 240Vac.  
Input Connector: V-M connector.  
Line Regulation: Typ. 0.1%.  
Load Regulation: Typ. ±1.0%.  
Noise & Ripple: Typ. 1.0% peak to peak.  
OVP: Built-in at main output (Latch).  
Adjustability: Factory Set.  
Input Frequency: 47-63Hz.  
Inrush Current: 12.0A rms cold start @230Vac.  
Input Current: Typ. 0.9A@115Vac, 0.435 @230Vac.  
Dielectric Withstand: Meet IEC 60601-1.  
EMI: Meet EN 55011 & FCC Class B.  
Hold-up Time: Typ. 11.2mS @115Vac, 72mS @ 230Vac.  
Earth Leakage: Less than 0.1mA @230Vac.  
Overload Protection (OLP):  
Fully protected against outputs overload and short circuit.  
Typ. OLP set at 105-120% of rating output power.  
Consult the factory for OLP setting.  
GENERAL SPECIFICATION  
Efficiency: Typ. 80-89.5 %. (Various with output voltage).  
Operating Temperature: 0 to +50 °C under forced air flow  
for full load. Derate by 1.3W/°C  
Switching Frequency: 85KHz.  
Circuit Topology: Fixed frequency flyback circuit.  
Transient Response: Output voltage returns in less than  
from 40°C to 50°C.  
Storage Temperature: -20 to +85°C.  
0.5mS following a 50% load change. Temperature Coefficient: Typ. 0.04%/°C.  
Safety Standard: Meet Medical standards UL 2601-1/  
EN 60601-1 Class I.  
Cooling: At least 5 cfm direct forward air flow is required to  
achieve full rating power.  
Power Density: 7.7Watts / Cubic Inch.  
Construction: 4”x2”x1” PCB format.  
Note: Due to requests in market and advances in technology, specifications subject to change without notification.  
For the details of safety approval, please consult the factory.  
-15-