AFE5808
www.ti.com
SLOS688C –SEPTEMBER 2010–REVISED APRIL 2012
Fully Integrated, 8-Channel Ultrasound Analog Front End with
Passive CW Mixer, 0.75nV/rtHz, 14/12-Bit, 65MSPS, 153mW/CH
Check for Samples: AFE5808
1
FEATURES
APPLICATIONS
•
8-Channel Complete Analog Front-End
LNA, VCAT, PGA, LPF, ADC, and CW Mixer
•
Medical Ultrasound Imaging
–
•
Nondestructive Evaluation Equipments
DESCRIPTION
•
Programmable Gain Low-Noise Amplifier
(LNA)
The AFE5808 is a highly integrated Analog Front-End
(AFE) solution specifically designed for ultrasound
systems in which high performance and small size
are required. The AFE5808 integrates a complete
–
–
–
–
24/18/12 dB Gain
0.25/0.5/1 VPP Linear Input Range
0.63/0.7/0.9 nV/rtHz Input Referred Noise
Programmable Active Termination
time-gain-control (TGC) imaging path and
a
continuous wave Doppler (CWD) path. It also enables
users to select one of various power/noise
combinations to optimize system performance.
Therefore, the AFE5808 is a suitable ultrasound
analog front end solution not only for high-end
systems, but also for portable ones.
•
40 dB Low Noise Voltage Controlled
Attenuator (VCAT)
•
•
24/30 dB Programmable Gain Amplifier (PGA)
3rd Order Linear Phase Low-Pass Filter (LPF)
The AFE5808 contains eight channels of voltage
controlled amplifier (VCA), 14/12-bit Analog-to-Digital
Converter (ADC), and CW mixer. The VCA includes
Low noise Amplifier(LNA), Voltage controlled
–
10, 15, 20, 30 MHz
•
•
14-bit Analog to Digital Converter (ADC)
–
–
77 dBFS SNR at 65 MSPS
LVDS Outputs
Attenuator(VCAT),
Programmable
Gain
Amplifier(PGA), and Low-Pass Filter (LPF). The LNA
gain is programmable to support 250 mVPP to 1 VPP
input signals. Programmable active termination is
also supported by the LNA. The ultra-low noise VCAT
provides an attenuation control range of 40dB and
improves overall low gain SNR which benefits
harmonic imaging and near field imaging. The PGA
provides gain options of 24 dB and 30 dB. Before the
ADC, a LPF can be configured as 10 MHz, 15 MHz,
20 MHz or 30 MHz to support ultrasound applications
with different frequencies. The high-performance 14
bit/65 MSPS ADC in the AFE5808 achieves 77 dBFS
SNR. It ensures excellent SNR at low chain gain. The
ADC’s LVDS outputs enable flexible system
integration desired for miniaturized systems.
Noise/Power Optimizations (Full Chain)
–
–
–
153 mW/CH at 0.75 nV/rtHz, 65 MSPS
98 mW/CH at 1.1 nV/rtHz, 40 MSPS
80 mW/CH at CW Mode
•
Excellent Device-to-Device Gain Matching
±0.5 dB(typical) and ±0.9 dB(max)
–
•
•
•
Low Harmonic Distortion
Fast and Consistent Overload Recovery
Passive Mixer for Continuous Wave
Doppler(CWD)
–
Low Close-in Phase Noise –156 dBc/Hz at 1
KHz off 2.5 MHz Carrier
–
–
–
–
Phase Resolution of 1/16λ
Support 16X, 8X, 4X and 1X CW Clocks
12dB Suppression on 3rd and 5th Harmonics
Flexible Input Clocks
•
Small Package: 15 mm x 9 mm, 135-BGA
NOTE
AFE5808A is an enhanced version of AFE5808 and it is recommended for new designs.
Compared to AFE5808, it expands the cut-off frequency range of the digital high pass
filter; increases the handling capability of extreme overload signals; lowers the correlated
noise significantly when high impedance source appears.
1
Please be aware that an important notice concerning availability, standard warranty, and use in critical applications of
Texas Instruments semiconductor products and disclaimers thereto appears at the end of this data sheet.
PRODUCTION DATA information is current as of publication date.
Products conform to specifications per the terms of the Texas
Instruments standard warranty. Production processing does not
necessarily include testing of all parameters.
Copyright © 2010–2012, Texas Instruments Incorporated