TRUMPower
130 WATT MEDICAL POWER ADAPTERS
TMP130 SERIES
DESCRIPTION
(MDD+EMC)
The TMP130 series of AC/DC switching power
adapters are for 100 to 130 watts of continuous
output power. They are designed for medical
applications that require external power. The units
are enclosed in a 94V-1 rated polyphenylene-oxide
case with an IEC 320 inlet to mate with
interchangeable cord for world-wide use. An on/off
switch can be added at request. All models meet
CISPR 11 and FCC class B emission limits, and
comply with UL, CSA, IEC and CE requirements.
FEATURES
‧ In compliance with EN61000-3-2, EN61000-3-3 and
EN61000-1-1-2
‧ Optional output connectors
‧ Optional on/off switch
‧ Optional Power Fail Detect (PFD) signal
‧ 100% burn-in at full load
Safety Standard Approvals:
‧ Auto-ranging input
UL 2601-1
CSA C22.2 No. 601.1
‧ Input surge current protection
‧ Overvoltage protection
‧
Thermal protection
INPUT SPECIFICATIONS
Input voltage:
90 to 132VAC or 180 to 264VAC
(auto-ranging)
Input frequency: 47 to 63 Hz
Input current: 3.0A (rms) for 115 VAC
1.7A (rms) for 230VAC
TÜV EN60601-1
Leakage current: 90µA max. @ 115VAC, 60Hz
150µA max. @ 230VAC, 50Hz
GENERAL SPECIFICATIONS
OUTPUT SPECIFICATIONS
Switching frequency: 68 Khz ±8Khz
Output voltage/current: See rating chart
Efficiency:
70% minimum on Vo ≥12V
Total output power:
Ripple and noise:
130 watts maximum
1% peak to peak maximum
65% minimum on others
Hold-up time:
Line regulation:
Inrush current:
10 msec minimum at 115VAC
±0.5% maximum at full load
Overvoltage protection: Provided on output: set at
112 – 132% of its nominal
15amps @115VAC, or 30amps
@230VAC, at 25ºC and cold start
4000VAC from input to output
1500VAC from input to ground
5000VAC from output to ground
400,000 hours minimum at full load
at 25ºC ambient, calculated per
MIL-HDBK-217F, excluding DC fan.
output voltage
Withstand voltage:
MTBF:
Overcurrent protection: All outputs protected to
short circuit conditions
Temperature coefficient: All outputs ±0.04%/ºC
maximum
EMC Performance (EN60601-1-2: 2001)
EN55011:
FCC:
VCCI:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
EN61000-4-3:
Class B conducted, class B radiated
Transient response:
Maximum excursion of 4% or
better on all models,
recovering to 1% of final
value within 500µs after a
25% step load change
Class B conducted, class B radiated
Class B conducted, class B radiated
Harmonic distortion, class A
Line flicker
ESD, ±8KV air and ±6KV contact
Radiated immunity, 3V/m for
80~2500MHz
Fast transient/burst, ±2KV
Surge, ±1KV diff, ±2KV com.
Conducted immunity, 3Vrms
Magnetic field immunity, 3A/m
Voltage dips, 30% reduction for
500ms, 60% reduction for 100ms and
>95% reduction for 10ms
ENVIRONMENTAL SPECIFICATIONS
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
EN61000-4-11:
0ºC to +70ºC
-40ºC to +85ºC
5% to 95% non-condensing
Derate from 100% at +50ºC
linearly to 50% at +70ºC
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com