TMP135 Series
120-135 Watts Medical & ITE PowerAdapters
TRUMPower
FEATURES
• 90~264 VAC Input, Soft Start
• 0.98 Active Power Factor
• 87% to 91% Efficiency
• High Peak Power
• Low Safety Ground Leakage Current
• Class I & Class II Models Available
• CEC & Energy Star Level V Compliance
INPUT SPECIFICATIONS
MODELS LIST
Output
Voltage
Maximum
Current
Peak (2)
Current
Maximum
Power
Input Voltage Range ............ 90~264 VAC
Input Frequency ................... 47~63Hz
Input Current ........................ 1.60A rms @ 115 VAC
0.80A rms @ 230 VAC
(1)
Product No.
TMP135-12
TMP135-13
TMP135-15
TMP135-18
TMP135-20
TMP135-24
TMP135-28
TMP135-30
TMP135-36
TMP135-48
12V
10.0A
9.23A
9.29A
7.50A
6.75A
5.63A
4.83A
4.50A
3.75A
2.94A
20.0A
18.5A
18.6A
15.0A
13.5A
11.3A
5.8A
120W
120W
130W
135W
135W
135W
135W
135W
135W
135W
13V
Inrush Current ..................... 80A @ 115 VAC or 160A @ 230 VAC,
14~16V
18~19V
20~21V
24~25V
28~29V
30~32V
36~38V
46~50V
°
at 25 C cold start
Soft Start............................... Yes
Earth Leakage Current ........ 200µA max. @ 264 VAC, 63Hz
Touch Current ...................... 100µA max. @ 264 VAC, 63Hz
OUTPUT SPECIFICATIONS
Output Power Ratings .......... See table
Minimum Load ...................... Not required
Tolerance ............................. 5%
Ripple and Noise* ............... 1% peak to peak max. at full load
Overvoltage Protection ........ Set at 115-140% of its nominal
output voltage
5.4A
4.5A
3.5A
Notes:
1) The listed PꢀN's are for Class I models equipped with IEC 320ꢀC14 AC inlet.
Add suffix "C2" to order Class II models equipped with IEC 320ꢀC18 AC inlet.
When ordering 14-50V output units, specify the requested voltage, e.g. use PꢀN
TMP135-18.5-C2 to order 18.5V Class II model.
2) The peak current is for 10 seconds maximum with the average power not over the
maximum power.
Overcurrent Protection ........ Protected to short-circuit
conditions
°
Temperature Coefficient ..... 0.04%ꢀ C max.
Transient Response........... Max. excursion of 4% or better on
all models, recovering to 1% of
final value within 500 us after a
25% step load change
STANDARDS & COMPLIANCES
EN 55011, EN 55022 ........ Class B, conducted & radiated
FCC, VCCI ......................... Class B, conducted & radiated
EN 61000-3-2.................... Harmonic distortion, Class A & D
EN 61000-3-3.................... Line flicker
* Measured with 20MHz bandwidth at rated line voltage and output load ranges, with
a 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor in parallel
across the output.
EN 61000-4-2.................... ESD, 8 KV air and 6 KV contact
EN 61000-4-3.................... Radiated immunity, 3Vꢀm
EN 61000-4-4.................... Fast transientꢀburst, 2 KV
EN 61000-4-5.................... Surge, 1 KV diff., 2 KV com.
EN 61000-4-6.................... Conducted immunity, 3 Vrms
EN 61000-4-8.................... Magnetic field immunity, 3Aꢀm
EN 61000-4-11 .................. Voltage dip immunity,
30% reduction for 500ms,
GENERALSPECIFICATIONS
Power Factor ....................... 0.98 typical @ 115 VAC
Efficiency ............................. 87% minimum at full load
Hold-up Time....................... 15ms at 110 VAC
Switching Frequency ........... 90~160KHz
°
°
Operating Temperature ...... 0 C to +60 C
Derating............................... Derate from 100% at +40 C linearly
°
°
to 50% at +60 C
60% reduction for 100ms,
>95% reduction for 10ms
°
°
Storage Temperature .......... -40 C to +85 C
°
°
Relative Humidity ................ 5 C to +95 C non-condensing
Withstand Voltage ............... 4000 VAC from input to output
1500 VAC from input to ground
Safety Standards ............... ULꢀIECꢀEN 60601-1 (3rd Edition),
ANSIꢀAAMI ES 60601-1 (1st Edition),
CSA C22.2 No. 60601-1 (2nd Edition),
500 VAC from output to ground
IECꢀEN 60950-1 (2nd Edition)
MTBF .................................... 150K hours minimum at full load,
Agency Approvals .............. UL, cUL, TUV, CE, CB
Other Compliances .......... RoHS, CEC & Energy Star Level V
°
25 C ambient, calculated per
MIL-HDBK-217F
* Class I models certified to medical and ITE safety standards; Class II models
certified to medical standard only.
Tumbler Technologies + TRUMPower