50/80/120 Watts
PCM Series
xppower.com
•
•
•
•
•
•
•
4000 VAC Isloation
Low Leakage Current
Single, Dual and Triple Outputs
Worldwide Medical Approvals
CCC Approved (PCM80)
Both Line and Neutral Fused
Non-standard Connectors Available
Specification
Input
General
Efficiency
• Single output models: 80ꢀ typical
Multi output models: 75ꢀ typical
Input Voltage
Input Frequency
Input Current
• 90-264 VAC
• 47-63 Hz
Isolation
• 4000 VAC Input to Output
1500 VAC Input to Ground
500 VAC Output to Ground
• PCM50:
PCM80:
1.35 A max
1.30 A max
PCM120: 1.60 A max
Switching Frequency
• PCM50: 64 kHz typical
PCM80: 55 kHz typical
PCM120: 120 kHz typical
Inrush Current
Power Factor
• 30 A at 115 VAC or
60 A at 230 VAC at 25 ˚C cold start
• PCM50: 0.55 typical
PCM80/120: 0.94 min
MTBF
• 100 kHrs at full load and 25 ˚C ambient
per MIL-HDBK-217F
Earth Leakage Current • 180 µA max at 132 VAC/60Hz
300 µA max at 264 VAC/50Hz
Environmental
Output
Operating Temperature • PCM50: 0 ˚C to +70 ˚C, derate from 100ꢀ
load at 50 ˚C to 50ꢀ load at +70 ˚C
Output Voltage
Minimum Load
• See tables
PCM80: 0 ˚C to +70 ˚C, derate from 100ꢀ
load at 40 ˚C to 50ꢀ load at +70 ˚C
PCM120: 0 ˚C to +60 ˚C, derate from
• No minimum load required for single
output models. See tables for multi
output models
100ꢀ load at 40 ˚C to 50ꢀ load at +60 ˚C
Start Up Delay
Hold Up Time
Line Regulation
• 2 s max
Storage Temperature
• PCM50/80: -40 ˚C to +80 ˚C
PCM120: -20 ˚C to +85 ˚C
• 10 ms min at low line
•
1ꢀ max at full load for 10ꢀ cꢁange
in input
EMC & Safety
Load Regulation
•
5ꢀ max
Emissions
• EN55022, level B conducted
EN55022, level B radiated
Transient Response
• 5ꢀ max. deviation, 500 µs recovery time
for a 50ꢀ load cꢁange
Harmonic Currents
Voltage Flicker
ESD Immunity
• EN61000-3-2, class A
Ripple & Noise
• 2ꢀ pk-pk, 20MHz BW
• EN61000-3-3 amendments 1 & 2
Overvoltage Protection • 112-132ꢀ witꢁ auto recovery on V1
• IEC61000-4-2, level 3
Perf Criteria B
Overcurrent Protection • 110-200ꢀ witꢁ auto recovery
on all outputs
Radiated Immunity
EFT/Burst
• IEC61000-4-3 10 V/m, level 3
Perf Criteria B
Temperature
Coefficient
•
0.05ꢀ /˚C
• IEC61000-4-4, level 3
Perf Criteria B
Surge
• IEC61000-4-5, level 3
Perf Criteria B
Conducted Immunity
• IEC61000-4-6 10 V/m, level 3
Perf Criteria B
Power Frequency
Magnetic Fields
• IEC61000-4-8 10 A/m, Perf Criteria B
Dips & Interruptions
• IEC61000-4-11, 30ꢀ 500 ms,
60ꢀ 100 ms, 100 ꢀ 5000 ms
Perf Criteria A, A, C
Safety Approvals
• UL60601-1, CSA C22.2 No. 601.1 per cUL
EN60601-1, CCC (PCM80)