Metal package PMT
Photosensor Modules H5784 Series
The H5784 series photosensor modules are comprised of a metal package pho-
tomultiplier tube, a low-power consumption high-voltage power supply and a low
noise amplifier. The electrical current from the photomultiplier tube is converted to
a voltage by an amplifier for easy signal processing. The H5784 is highly resistant
to noise since the amplifier is installed near the anode output pin of the photomul-
tiplier tube. The amplifier feedback resistance of 1 MΩ allows a current-to-voltage
conversion factor of 1 V/µA, and covers a frequency bandwidth from DC to 20
kHz.
Product Variations
Current-to-voltage Frequency
Type No. Spectral Response
Features
Conversion Factor Bandwidth
H5784
300 nm to 650 nm
For general applications in visible range
For general applications in visible to near IR range
High sensitivity in near IR range.
H5784-01 300 nm to 850 nm
H5784-02 300 nm to 880 nm
H5784-03 185 nm to 650 nm
H5784-04 185 nm to 850 nm
H5784-06 185 nm to 650 nm
H5784-20 300 nm to 900 nm
1 V/µA
DC to 20 kHz For UV to visible range
For UV to near IR range
For UV to visible range (synthetic silica window) with higher sensitivity below 300 nm than -03 type
Infrared-extended multialkali photocathode with enhanced sensitivity
Specifications
Unit
—
V
H5784 Series
-01/-04
Parameter
Suffix
None/-03/-06
-02
-20
Input Voltage
11.5 to 15.5
18
Max. Input Voltage
Max. Input Current
Max. Output Signal Voltage
Max. Control Voltage
V
+9/-1
mA
V
+10 (load resistance 10 kΩ)
+1.0 (Input impedance 100 kΩ)
V
Recommended Control Voltage Adjustment Range
Effective Area
+0.25 to +0.9
V
8
mm
—
nm
Sensitivity Adjustment Range
Peak Sensitivity Wavelength
1: 104
420
400
80
500
200
630
350
Min.
40
Luminous Sensitivity
µA/lm
Typ.
Blue Sensitivity Index (CS 5-58)
Red/White Ratio
70
150
—
250
500
8
—
—
—
—
—
0.2
0.25
0.45
Radiant Sensitivity *1
62
60
58
78
mA/W
Min.
1.0 × 107
5.0 × 107
43
1.5 × 107
7.5 × 107
30
2.5 × 107
1.25 × 108
29
3.5 × 107
2.5 × 108
39
2
Luminous Sensitivity
*
V/lm
V/nW
mV
Typ.
Radiant Sensitivity *1 *2
Voltage Output Depending
on PMT Dark Current *2 *3
Typ.
0.2
0.4
2
2
Max.
2
4
20
20
Current-to-Voltage Conversion Factor
Offset Voltage *2
Typ.
Ripple Noise *2 *4 (peak to peak) Max.
Settling Time *5
1
V/µA
mV
mV
s
3
2
2
Operating Ambient Temperature
Storage Temperature
+5 to +50
°C
°C
g
-20 to +50
100
Weight
*1: Measured at the peak sensitivity wavelength
*2: Control voltage = +0.8 V
*3: After 30 minute storage in darkness
*4: Cable RG-174/U, Cable length 450 mm, Load resistance = 1 MΩ, Load capacitance = 22 pF
*5: The time required for the output to reach a stable level following a change in the control voltage from +1.0 V to +0.5 V.
28