GPC200 Commercial/GPM200 Medical
200 Watt Global Performance Switchers
FEATURES:
• Wide-range ac input 85-264 Vac
• 2-year warranty
• Conducted EMI exceeds FCC Class B and CISPR 22
Class B (Commercial models) and CISPR 11 Class B
(Medical models)
• Commercial Approved to UL1950, IEC950 and
CSA22.2-234 L3
• Medical Approved to UL2601-1, IEC601-1 and
CSA22.2 No. 601
• Multiple outputs
•
marked to LVD
Power Fail
SPECIFICATIONS:
TTL- or CMOS-compatible output goes low (< 0.5V) 5 ms
before output voltage drops more than 4% below nominal
voltage upon loss of ac power.The signal is factory set to trip on
84 to 94Vac brown-out depending upon incoming line impedance
and distortion. Other settings are available to the user through
adjustment of built-in potentiometer.
Ac Input
85-264 Vac, 47-63 Hz single phase.
Input Current
Maximum input current at 120Vac, 60 Hz with full rated output
load: 5.5 A
Output Noise
Hold-UpTime
0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential mode.
Measured with noise probe directly across output terminals of
the power supply.
20 ms minimum from loss of ac input at full load, nominal line
(115Vac).
Output Power
Transient Response
150 W convection;200 with 26 cfm air;220 W peak. Peak ratings
are for 60 s maximum duration, 10% duty cycle. During peak
load condition, output regulation may exceed total regulation
limits.
Main output—500 µs typical response time for return to within
0.5% of final value for a 50% load step change. ∆i/∆t<0.2 A/µs.
Maximum voltage deviation is 3.5%. Startup/shutdown overshoot
less than 3%.
Overload Protection
Remote Sense
Fully protected against short circuit and output overload. Short
circuit protection is cycling type power limit on outputs 1 & 2;
foldback type on outputs 3 & 4. Recovery after fault is automatic.
See output ratings chart for additional notes or conditions.
Provided as a standard feature on main 5V output.
Voltage Adjustment
Built-in potentiometer adjusts voltage ±5% on output 1 & 2.
Overvoltage Protection
EMI/EMC Compliance
Main outputs: 124% ± 12% typical.
All models include built-in EMI filtering to meet the following
emissions requirements:
Efficiency
EMI SPECIFICATIONS
Conducted Emissions GPC200 EN55022 Class B; FCC Class B
Conducted Emissions GPM200 EN55011 Class B; FCC Class B
COMPLIANCE LEVEL
Minimum 65% at full rated load, nominal input voltage, depending
on model and load distribution.
Static Discharge
EN61000-4-2, 6 kV contact, 8 kV air
EN61000-4-3, 3 V/meter
EN61000-4-4, 2 kV, 5 kHz
EN61000-4-5, 1 kV diff., 2 kV com.
Input Protection
Internal ac fuse provided. Designed to blow only if a catastrophic
failure occurs in the unit.
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Inrush Current
Commercial Leakage Current
Inrush is limited by internal thermistors.Inrush at 240Vac under
cold start conditions will not exceed 60 A.
0.95 mA 254Vac @ 60 Hz input (with no deviations).
Commercial Safety
Temperature Coefficient
0.03%/°C typical on all outputs.
Approved to UL1950, CSA22.2 No. 234 Level 3, IEC950 and
EN60950. UL file #E135803 commercial; CSA #LR46516 all
models.The output(s) are intended for safety earthed Signal
Output and Intermediate Circuits only.All dc outputs are SELV
under normal and single fault conditions.
Thermal Shutdown
Provided as a standard feature. Designed to protect unit from
prolonged overtemperature.
Medical Leakage Current
60 µA 254 Vac @ 60 Hz input.
Environmental
Designed for 0 to 50°C operation at full rated output power;
derate output current and total output power by 2.5% per °C
above 50°C. See Environmental and Packaging Specifications
on next page.
Medical Safety
Approved to UL2601-1, CSA-C22.2 No. 601 Level 3 and IEC601.
UL file E116994;CSA #LR46516.The output(s) are intended for
safety earthed Signal Output and Intermediate Circuits only.The
output(s) are not acceptable for patient connection without
additional isolation.All dc outputs are SELV under normal and
single fault conditions.