AFE49I30
SBASA49 –MAY 2020
AFE49I30 Ultra-Small, Integrated AFE With FIFO for Multi-Sensor Wearable, Optical Heart-
Rate Monitoring and Bio-sensing
1 Features
•
Supports external clock and internal oscillator
modes
1
•
•
Supports signal acquisition of up to 24 phases
•
Option to acquire data synchronized with a system
master clock
Flexible allocation of 6 LED, 4 PD in each PPG
phase
•
•
•
•
•
Automatic cancellation of DC from ambient, LED
FIFO with 128-sample depth
•
•
Synchronized PPG/ECG signal acquisition with
each phase configurable as PPG or ECG, and at
different data rates
I2C interface
Synchronized PPG/ECG supports up to 1 kHz
data rate
2.6-mm × 2.1-mm DSBGA, 0.4-mm pitch
Supplies: Rx: 1.7 V-1.9 V (LDO bypass); 1.9 V-3.6
V (LDO enabled), Tx: 3 V-5.5 V, IO: 1.7-3.6 V
•
•
ECG-only mode supports up to 4 kHz data rate
PPG signal chain:
•
Supports operation with 2 supplies - 1.8 V, 5 V
–
–
System SNR up to 108-dB at 16 μA PD current
2 Applications
Low current for continuous operation on a
wearable device with a typical value: 15 μA for
an LED, 20 μA for the receiver
•
•
ECG on a wearable device
Synchronized PPG, ECG for PTT-based Blood
pressure
•
PPG Transmitter:
•
Optical heart-rate monitoring (HRM) and Heart-
rate variability (HRV) for wearables, hearables
–
–
Supports common anode LED configuration
8-Bit programmable LED current with range
adjustable from 25 mA to 250 mA
•
•
Pulse oximetry (SpO2) measurements
Maximum oxygen consumption (VO2 Max)
–
Mode to fire two LEDs in parallel with
independent per-phase current control
3 Description
–
–
Programmable LED on-time per-phase
The AFE49I30 is an analog front-end for optical bio-
sensing/ PPG and ECG measurement applications.
The device supports up to six switching light-emitting
diodes (LEDs) and up to four photodiodes. Up to 24
signal phases can be defined and the PPG or ECG
signals can be acquired from each phase in a
synchronized manner.
Simultaneous support of 6 LEDs for SpO2,
HRM, or multi-wavelength HRM
•
PPG Receiver:
–
–
Supports 4 time-multiplexed photodiode inputs
2 parallel receivers to sample two PD channels
simultaneously
Device Information(1)
–
–
Individual ambient offset subtraction DAC at
each TIA Input with 8-bit per-phase control and
range adjustable up to 255-μA
PART NUMBER
PACKAGE
BODY SIZE (NOM)
AFE49I30
DSBGA (30)
2.60 mm × 2.10 mm
Individual LED offset subtraction DAC at each
TIA input with 5-bit per-phase control and
15.5-µA range
(1) For all available packages, see the orderable addendum at
the end of the data sheet.
Simplified Schematic
–
–
–
Close to 100 dB ambient rejection up to 10 Hz
Noise filtering with programmable bandwidth
Transimpedance Gain: 3.7 kΩ to 1 MΩ
•
ECG Signal Chain
–
–
–
1-lead ECG with RLD output for bias
INA gain programmable from 1.6 – 13
<1 µVrms input noise (in a 1-150 Hz
bandwidth) at 4-kHz sample rate
–
–
Integrated 450 Hz Low Pass Filter
AC, DC lead-off detect with 3.125-100 nA
current and programmable interrupt
1
An IMPORTANT NOTICE at the end of this data sheet addresses availability, warranty, changes, use in safety-critical applications,
intellectual property matters and other important disclaimers. PRODUCTION DATA.