WMIHPU15 SERIES
Up to 15 Watts, Medical AC/DC Wall Mount Power Supplies
Rev A
Wall Industries, Inc.
FEATURES
Class II System
RoHS Compliant
Single Outputs Ranging from 3VDC to 55VDC
-10°C to +70°C Operating Temperature Range
Up to 15 Watts Output Power
86% High Efficiency
Meets FCC Part-18 Class B & CISPR-11 EN55011
Class B Emission Limits
Energy Star 2.0, Efficiency Level V
80-275VAC Input Voltage Range
Over Voltage & Over Load Protection
Input to Output: 2MOPP
UL/cUL 3rd Edition Medical Approvals
Interchangeable Plug Options: United States,
Europe, Australia, & United Kingdom Types
Optional Output Connectors Available
SAFETY APPROVALS
Size:
Weight:
5.82oz (165g)
2.36 x 1.71 x 1.58 inches
60.0 x 43.5 x 40.2 mm
DESCRIPTION
The WMIHPU15 series of Class II medical AC/DC wall mount power supplies offers up to 15
watts of output power in a 2.36” x 1.71” x 1.58” package. This series consists of single output
models ranging from 3VDC to 55VDC with an 80~275VAC input voltage range and a -10°C
to +70°C operating temperature. This series meets FCC Part-18 Class B and CISPR-11
EN55011 Class B Emission Limits and has ANSI/AAMI ES 60601-1: 2005 (UL/cUL 3rd edition)
and CE safety approvals. All units are RoHS and Energy Star Level V compliant. Plugs come
in United States (“U” suffix), Europe (“E” suffix), Australia (“A” suffix), and United Kingdom
(“K” suffix) types. Plugs are sold separately so please contact factory for ordering details.
AC Plug Types:
Applications:
• Home Healthcare Equipment
• Patient Monitors
• Unites States (“U” Suffix)
• Europe (“E” Suffix)
• Australia (“A” Suffix”)
• United Kingdom (“K” Suffix) • Portable Medical Devices
• ECG Machines
• Blood Pressure Systems
MODEL SELECTION TABLE
Adjustable Voltage Range Adjustable Current Range
Maximum
Output Power
Total
Regulation
No Load
Consumption
Model Number (1) Input Voltage
Ripple & Noise
Efficiency
Min
Max
Min
Max
WMIHPU15-101x
WMIHPU15-102x
WMIHPU15-103x
WMIHPU15-104x
WMIHPU15-105x
3.0 VDC
4.99 VDC
1.44 A
2.40 A
7.2W
12W
12W
13.5W
15W
15W
15W
15W
15W
15W
15W
15W
66mVp-p
50mVp-p
70mVp-p
90mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
5%
5%
5%
5%
5%
5%
5%
5%
3%
3%
3%
3%
65%
80%
82%
83%
84%
85%
85%
86%
86%
86%
86%
86%
0.2W
0.3W
0.3W
0.3W
0.3W
0.3W
0.3W
0.3W
0.3W
0.3W
0.3W
0.3W
5.0 VDC
6.5 VDC
5.99 VDC
8.0 VDC
2.00 A
1.50 A
1.22 A
1.15 A
0.94 A
0.72 A
0.55 A
0.45 A
0.37 A
0.32 A
0.28 A
2.40 A
1.84 A
1.68 A
1.36 A
1.15 A
0.94 A
0.72 A
0.55 A
0.45 A
0.37 A
0.32 A
8.0 VDC
11.0 VDC
13.0 VDC
16.0 VDC
21.0 VDC
27.0 VDC
33.0 VDC
40.0 VDC
48.0 VDC
55.0 VDC
11.0 VDC
13.0 VDC
16.0 VDC
21.0 VDC
27.0 VDC
33.0 VDC
40.0 VDC
48.0 VDC
WMIHPU15-106x
80 ~ 275VAC
WMIHPU15-107x
WMIHPU15-108x
WMIHPU15-109x
WMIHPU15-110x
WMIHPU15-111x
WMIHPU15-112x
NOTES
1. The “x” in the model number can be “U” for United States type plug; “E” for Europe type plug, “A” for Australia type plug, or “K” for United Kingdom type
plug. Plugs are sold separately so please contact factory for ordering details.
2. Output can provide up to peak load when power supply starts up. Continuous staying in more than rated load is not allowed.
3. At factory, in 60% rated load condition, each output is checked to be within voltage accuracy.
4. Line regulation is defined by changing 10% of input voltage from nominal line at rated load.
5. Load regulation is defined by changing 40% of measured output load from 60% rated load.
6. Ripple & noise is measured by using 20MHz limited bandwidth and with 0.47F capacitor in parallel across the output at nominal line and rated load.
7. Hold-up time is measured from the end of the last charging pulse to the time which the main output drops down to low limit of main output at rated
load and nominal line.
8. Optional output connectors are available for this series. Please call factory for ordering details.
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
08/04/2014
Page 1 of 4
website: www.wallindustries.com • e-mail: sales@wallindustries.com