THURLBY THANDAR INSTRUMENTS
TSX series High Current Laboratory Power Supplies More Info>
METER SPECIFICATIONS
Meter types:
Separate 4 digit meters for voltage and current with
12.5mm (0.5") LED displays.
10mV, 10mA.
Meter resolutions:
Meter accuracies:
Voltage ±(0.2% + 1 digit)
Current ±(0.5% + 1 digit).
MECHANICAL & ENVIRONMENTAL
Electrical safety:
EMC:
Temperature:
Complies with EN61010-1.
Complies with EN50081-1 and EN50082-1.
+5oC to +40oC operating, 20% to 80% RH, -40oC to
+70oC storage.
Size:
210 x 130 x 350mm (WxHxD)
(half rack width x 3U height), optional rack mounting
kit available.
● High power levels in a compact & lightweight casing
● 35V-10A & 18V-20A models with more to come
● Bench or rack mounting, front & rear terminals
● Very low noise, excellent transient reponse
Weight:
5.0kg (TSX versions).
5.5kg (TSX-P versions).
FRONT PANEL CONTROLS (standard versions)
Voltage setting:
Current setting:
Overvoltage setting:
Output On/Off:
Via single rotary controls for coarse and fine control.
Via single turn semi-logarithmic rotary control.
Via screwdriver adjustable preset potentiometer.
Via large paddle lever switch.
● CV & CI operation with automatic crossover
● Comprehensive protection including variable OVP trip
● High setting resolution, remote sense terminals
● High accuracy digital meters, current meter damping
● Keyboard/Rotary setting of all parameters (TSX-P)
● Watts display, non-volatile storage of 25 settings (TSX-P)
● Fully programmable with bus readback of V and I (TSX-P)
● GPIB (.2) & RS232 interfaces (TSX-P)
FRONT PANEL CONTROLS (P versions)
Voltage setting:
Direct keyboard entry or quasi-analogue
rotary control.
Direct keyboard entry or quasi-analogue
rotary control.
Direct keyboard entry.
Push button with dual indicator lamps.
Current setting:
Overvoltage setting:
Output On/Off:
Note: all voltage and current levels set via the keyboard are displayed on
a separate 0.3" 4 digit display. This entry preview system ensures that
the user can observe the value entered before it is effected thus avoiding
possible error. The display is also used for setting additional functions
and for displaying watts.
Model Range:
TSX1820 - Single output, 0 to 18V at 0 to 20A.
TSX3510 - Single output, 0 to 35V at 0 to 10A.
TSX1820P - 0 to 18V at 0 to 20A, GPIB & RS-232 interfaces.
TSX3510P - 0 to 35V at 0 to 10A, GPIB & RS-232 interfaces.
Additional keyboard
functions:
Increase or decrease voltage or current in
user-selectable steps (delta mode).
Store/recall voltage, current & OVP levels from
non-volatile memory (25 memories).
Set digital interface type (RS232 or GPIB), set baud
rate, set address.
OUTPUT SPECIFICATIONS
Operating modes:
Voltage range:
Current range:
Constant voltage or constant current with automatic
crossover.
0V to 35V (TSX3510/TSX3510P).
0V to 18V (TSX1820/TSX1820P).
0A to 10A (TSX3510/TSX3510P).
0A to 20A (TSX1820/TSX1820P).
DIGITAL INTERFACES (P versions)
RS232:
Variable baud rate, 9600 baud maximum, 9 pin D
connector (male). Fully compatible with standard
RS232 or TTi addressable RS232 system (ARC).
Conforming with IEEE488.1 & IEEE488.2.
Overvoltage protection: 10% to 110% of max. output voltage.
IEEE-488 (GPIB):
Setting resolution:
Load regulation:
Line regulation:
10mV, 10mA.
Operational functions: Set voltage; set current; set OVP; set output On/Off;
read output voltage/current.
Setting resolution:
<0.01% of max. O/P for 90% change.
<0.01% of max. O/P for 10% change.
<1mΩ in constant voltage mode.
>5kΩ in constant current mode.
<1mV RMS typical in constant voltage.
<3mA RMS typical in constant current.
Voltage - 10mV; Current - 10mA.
Voltage - ±(0.1% + 10mV);
Current - ±(0.2% + 20mA).
Interface - <15ms (single command);
PSU - Depends on Load conditions, typically 150ms
to within 0.1% of final value (except for voltage re-
duction with low load current which will be longer).
Voltage - 10mV; Current - 10mA.
Voltage - ±(0.1% + 1 digit);
Current - ±(0.5% +1 digit).
Software for operating the PSUs under GPIB or
RS232 control is available including a Labwindows*
driver and ARC-TALK software for a PC.
Output impedance:
Setting accuracy:
Response times:
Ripple & noise:
HF common mode noise:Typically <3mV RMS, <10mV pk.
Transient load response:<20us to within 50mV of set level for 90% load
change.
Readback resolution:
Readback accuracy:
Temperature coefficient: typically <100ppm/oC.
Overvoltage protection
Operating software:
delay:
<200us.
Protection functions:
Overvoltage trip, Regulator overtemperature
Sense miswiring.
Status indication:
Output on/off lamp, Constant voltage mode lamp
Constant current mode lamp, Trip message.
Electronic.
Thurlby Thandar Instruments Ltd. operates a policy of continuous development
and reserves the right to alter specifications without prior notice.
Output switch:
Output terminals:
4mm output terminals at front, screw terminals for
output and sense at rear.
Designed and built in the EEC by:
Output protection:
Full forward and reverse protection via OVP and di-
ode clamp.
INPUT SPECIFICATIONS
Input voltage range:
Power requirement:
Voltage range selection: Rear panel slide switch.
180V to 270V RMS, 90V to 135V RMS, 47 to 63Hz.
750VA max.
Thurlby Thandar Instruments Ltd.
Glebe Road, Huntingdon. Cambs. PE18 7DX England
Tel: +44 (0)1480 412451 Fax: +44 (0)1480 450409
e-mail: sales@ttinst.co.uk Web: http://www.ttinst.co.uk
Note: This is a faxable data sheet, a colour brochure is also available.