SDT310
small type platinum thin film thermal sensors
NEW
features
• T.C.R. is in accordance with JIS-DIN standards
• The small package with a real ability of 1kΩ resistance
• Thermal time constant is improved with the small package
• Products with lead-free terminations meet RoHS
requirements. RoHS regulation is not intended for
Pb-glass contained in glass
Dimensions inches (mm)
dimensions
Type
L
W
t
f
d (Nom.)
l
.008±.002
( 0.2±0.05)
and construction
+.179
SDT310LT
.394
-.079
Lead Weld
(10 +5
)
Reinforcing Coating
L
l
-2
SDT310LTC
SDT310
Protective
Coating
Lead Wire
.118±.010 .079±.010 .047 max. .043±.010
(3.0±0.25) (2.0±0.25) (1.2 max.) (1.1±0.25)
.591±.179
(15±5)
t
W
.008±.002
(ø0.2±0.05)
f
SDT310MT
SDT310N
Welding
Point
.315±.079
(8±2)
Platinum
Thin Film
Ceramic
Substrate
.157±.004 .063±.004 .039 max. .031±.010
(4.0±0.1) (1.6±0.1) (1.0 max.) (0.8±0.25)
ordering information
New Part #
SDT310
2B
LT
C
100
B
3850
Temperature
Range
Terminal Surface
Temperature
Nominal
Resistance
T.C.R.
Type
Size Code
Class
(x 10-6/K)
N: 1.6x4.0mm
Nil: 2.0x3.0mm
LT: -55°C~+155°C
C: SnCu
(SDT310LT)
100: 100Ω
500: 500Ω
1K: 1kΩ
B: ±(0.3+0.005[t])
C: ±(1.0+0.01[t])
Nil: -55°C~+400°C
MT:-55°C~+650°C
P: Pt clad wire
(SDT310-310N)
M: PtIr
(SDT310MT)
applications and ratings
Class:
Tolerance of
Measuring
Resistance
Range
@ 0°C
Thermal
Time
Constant***
Thermal
Dissipation
Constant***
Operating
Temperature
Range
Part
Designation
Resistance
Tolerance
T.C.R.
(ppm/°C)**
Specified
Current****
Temperature*
B:±(0.3+0.005 [t])
C:±(1.0+0.01 [t])
±12%
±39%
SDT310LT 100Ω, 500Ω,
-55°C to
+155°C
SDT310LTC
1kΩ
B:±(0.3+0.005 [t])
C:±(1.0+0.01 [t])
B:±(0.3+0.005 [t])
C:±(1.0+0.01 [t])
B:±(0.3+0.005 [t])
C:±(1.0+0.01 [t])
±12%
±39%
±12%
±39%
±12%
±39%
7 seconds in
stationary air
100Ω, 500Ω,
1kΩ
-55°C to
+400°C
SDT310
0.9mW/°C
1.0mW/°C
3850
1mA max.
-55°C to
+650°C
SDT310MT
SDT310N
100Ω
100Ω
9 seconds in
stationary air
-55°C to
+400°C
*
[t] is a measuring temperature indicated at °C that is not related to marking + or -.
** T.C.R. measuring temperature: o°C/+100°C.
*** Thermal time constant and thermal dissipation constant are values measured in stationary air and are typical values,
which are values of elements and vary with connecting or fixing methods.
**** Specified current is a current value that is used at reliability test under the condition of self heat-generation that can be disregarded.
Recommended measuring currents are 1mA for 100Ω and 0.1mA for 500Ωor 1kΩ.
For further information on packaging, please refer to Appendix A.
Specifications given herein may be changed at any time without prior notice. Please confirm technical specifications before you order and/or use.
1/13/06
156