PSMBU152 SERIES
150W Medical AC/DC Open Frame Power Supplies
Rev A
Wall Industries, Inc.
FEATURES
Class I
Up to 89% High Efficiency
RoHS Compliant
Active Power Factor Correction
Wide Input Voltage Range: 90~260VAC, 47~63Hz
Over Voltage & Over Current Protection
5” x 3” x 1.44” Open Frame Package
12VDC & 24VDC Single Outputs
150 Watts Output Power
Meets FCC Part-18 Class B & CISPR-11 EN55011 Class B Emission Limits
ANSI/AAMI ES 60601-1: 2005 (UL/cUL 3rd Edition) & EN 60601-1: 2006
(TUV/T-mark 3rd Edition)
Input to Output: 2MOPP
100% Burn-in Tested
DESCRIPTION
The PSMBU152 series of class I medical AC/DC switching power supplies provides 150 Watts of continuous output power in
a 5.00” x 3.00” x 1.44” open frame package. This series consists of 12VDC and 24VDC single output models with a wide input
voltage range of 90~260VAC. Some features include high efficiency up to 89%, active power factor correction, 2MOPP
insulation, and over current and over voltage protection. All models meet FCC Part-18 class B and CISPR-11 EN55011 class B
emission limits. This series also has ANSI/AAMI ES 60601-1: 2005 (UL/cUL 3rd edition) and EN 60601-1: 2006 (TUV/T-mark 3rd
edition) medical approvals. All models are RoHS compliant and have been 100% burn-in tested.
Size:
5.00 x 3.00 x 1.44 inches
127.0 x 76.2 x 36.6 mm
SPECIFICATIONS: PSMBU152 SERIES
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Typ
Max
Unit
INPUT SPECIFICATIONS
Safety Approvals Input Voltage Range
Operating Input Voltage Range
100
90
47
1.74
0.71
240
260
63
1.76
0.73
50
100
1.2
1
Input Voltage
Input Frequency
Input Current
VAC
Hz
A
100VAC, full load
240VAC, full load
115VAC, full load, 25°C, cold start
230VAC, full load, 25°C, cold start
230VAC, no load
Inrush Current
A
No Load Power Consumption
Power Factor Correction
OUTPUT SPECIFICATIONS
Output Voltage
W
240VAC, full load
0.95
See Table
Line Regulation
Load Regulation
Output Power
LL to HL, full load
230VAC
0.5
3
1
5
150
%
%
W
Output Current
See Table
Ripple & Noise (peak to peak)
Hold-up Time
Start-up Time
Transient Response Time
Temperature Coefficient
PROTECTION
90VAC, full load
110VAC, full load
100VAC, full load
100VAC, Full load to half load
0~50°C
1
%
ms
s
ms
%/°C
20
2
4
-0.04
+0.04
Over Voltage Protection
Over Current Protection
GENERAL SPECIFICATIONS
Efficiency
112
110
132
150
%
%
230VAC, full load
Primary to Secondary (2MOPP Insulation)
Primary to PE
Test Voltage = 500VDC
240VAC/60Hz
84
6990
2121
50
89
%
Dielectric Withstanding Voltage
VDC
Isolation Resistance
Leakage Current
MΩ
mA
0.25
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Derating linearly from 100% Load at 50°C to 50% load at 70°C
0
-40
0
+70
+85
95
95
3000
°C
°C
%
%
m
0
Operating Altitude
Cooling
Free air convection
MTBF
MIL-HDBK-217F, 25°C
100,000
hours
PHYSICAL SPECIFICATIONS
Weight
14.82oz (420g)
Dimensions (L x W x H)
Input Connector
5.00 x 3.00 x 1.44 inches (127.0 x 76.2 x 36.6 mm)
See mechanical drawing
Output Connector
SAFETY & EMC
See mechanical drawing
Safety Approvals
ANSI/AAMI ES 60601-1: 2005 (UL/cUL 3rd edition); EN 60601-1:2006 (TUV/T-mark 3rd edition); CE
EMI Requirements for CISPR-11
EMI Requirements for FCC PART-18
220VAC
110VAC
B
B
Class
Class
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: sales@wallindustries.com
11/18/2013
Page 1 of 3