Features:
•
•
•
•
•
•
•
•
•
Low safety ground leakage current
Meet EN55011 and FCC Class B
Small size, light weigh
100% burn-in
Wide input range 85-264 VAC
Input surge current protecꢀon
Overvoltage protecꢀon
Overcurrent protecꢀon
Compliant with RoHS requirements
Descripꢀon:
The PM110 series of compact, open PCB constructed, AC-DC switching power supplies are specially designed for medical
applicaꢀons. They are capable of delivering 72-110 waꢁs of conꢀnuous power at 25 CFM forced air cooling or 60-80 waꢁs at
convecꢀon cooling. They operate at 85-264VAC input voltage without the need of a selector strap. All models meet the
safety requirements of UL, CSA and IEC for medical equipment.
Output #12
Imin Imax Tol. V2 Imin Imax Ipeak4 Tol.
Output #2
Output #3
Output #4
Max.
Model1
Output
V1
V3
Imin Imax Tol. V4 Imin Imax
Tol.
Power3
PM110-10-1A 3.3V
0A
0A
0A
0A
0A
0A
22A ±3%
22A ±3%
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
60/72W
80/110W
80/110W
80/110W
80/110W
80/110W
PM110-10A
PM110-12A
PM110-13A
PM110-14A
PM110-16A
5V
12V
15V
24V
30V
9A
±2%
7.5A ±2%
4.5A ±2%
3.6A ±3%
PM110-23A
PM110-31A
PM110-32A
PM110-40A
PM110-41A
PM110-42A
PM110-45A
5.1V
5.1V
5.1V
5.1V
5.1V
5.1V
5.1V
0A
0A
0A
0A
0A
0A
10A ±3% 12V 0A
10A ±3% 12V 0A
10A ±3% 15V 0A
10A ±3% 12V 0A
10A ±3% 15V 0A
10A ±3% 12V 0A
5A
5A
4A
5A
4A
5A
9A
9A
±3%
±3%
80/110W
80/110W
80/110W
80/110W
80/110W
80/110W
80/110W
-12V
0A
0A
0A
0A
0A
1A
1A
1A
1A
1A
±4%
±4%
±4%
(N/A)
(N/A)
7.5A
9A
±3% -15V
±3% -12V
±3% -15V
±3% -12V
-5V
0A
0A
0A
0A
1A
±4%
±4%
±4%
±4%
7.5A
9A
±4% 24V
±4% 12V
±4% 24V
1A
1A
0A
0A
0A
0A
10A ±3% 12V 0A
10A ±3% 12V 0A
10A ±3% 24V 0A
10A ±3% 15V 0A
5A
5A
3A
4A
9A
9A
±3% -12V
±3% -12V
±3% -12V
±3% -15V
0A
0A
0A
0A
1A
1A
1A
1A
1A
3A
1A
1A
PM110-45-1A 5.1V
PM110-45-2A 5.1V
±4% 24V 1.5A
±10% 80/110W
5A
7.5A
±4% 12V
0A
0A
±4%
±4%
80/110W
80/110W
PM110-46A
5.1V
±4% -5V
NOTES:
1. Safety agency approvals are for the above listed models in PCB format. To order a model with a metallic L-bracket or box, change suffix “A” to “B” for
L-bracket format, to “C” for enclosed form with cover, e.g. PM110-14C. (mechanical details shown in Annex H)
2. The output #1 of model PM110-45-1A needs a minimum current of 2A to support the other outputs at their maximum rated load.
3. 110 waꢁs maximum at 25 CFM forced air cooling or 80 waꢁs maximum at convecꢀon cooling, except model PM110-10-1A which is rated at 60 waꢁs
maximum at convecꢀon cooling or 72 waꢁs maximum at 25 CFM forced air cooling.
4. Peak output current with 10% maximum duty cycle for less than 60 seconds. Total peak power must not exceed 130 waꢁs.
5. All models may be operated at no-load. At no-load, output voltage tolerance increases to ±10%.
6. Ripple and noise is maximum peak to peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges,
and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across the output.
TT Electronics | Power Partners, Inc
General Note
43 Broad Street Suite B206, Hudson, MA 01749, USA. t: +1 (978) 567-9600
All data sheets are subject to change without noꢀce.
ISS.1 22/07/2020 Page: 1
© TT Electronics plc