PM110 Series
7 2 - 1 1 0 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L
Features
• Low safety ground leakage current.
• Wide input range 85 to 264Vac.
• Input surge current protection.
• Overvoltage protection.
• Overcurrent protection.
• Open PCB construction.
• 100% burn-in at full rated load.
OPERATING
Specifications
INPUT
MTBF
400,000 hrs @ 25ꢁC
66ꢀ–75ꢀ
Input Voltage
85–264Vac wide input
47Hz–67Hz
Efficiency
Frequency
ENVIRONMENTAL
Operating Temperature
Derating
Input Current
3.2A (rms) for 115Vac.
1.8A (rms) for 230Vac.
0ꢁC to 70ꢁC
Leakage current
Isolation
90µA max. @ 115Vac. 60Hz.
150µA max. @ 230Vac. 50Hz.
Derate to 50ꢀ at 70ꢁC
-40ꢁC to 85ꢁC
Storage Temperature
Relative Humidity
Input – Output: 4000VAC
Input – Ground: 1500VAC
Output – Ground: 500VAC
5ꢀ–95ꢀ non-condensing
STANDARDS AND APPROVALS
Inrush Current
OUTPUT
15A @ 115V, 30A @ 230V
Safety standards
C-Tick
UL2601-1, CSA C22.2, TUV EN60601-1
AS/NZS CISPR11 Group 1, Class A
Output Voltage
See table
EMC standards
EN60601-1-2: 2001,EN61000-3-2, -3,
EN61000-4-2, -3, -4, -5, -6, -8, -11
Output Voltage Adjustment V1 10ꢀ
EMI standards
MECHANICAL
Dimensions
EN55011 Class B, FCC Class B, VCCI Class B
Output Current
Ripple & Noise
Line Regulation
See table
1ꢀ peak to peak
0.5ꢀ maximum at full load.
See table
177.8x108x47mm
640grams.
Weight
Load Regulation
Transient Response
Max. deviation 4ꢀ recovering to 1ꢀ within
500µs after 25ꢀ load charge
Overvoltage Protection
Provided on V1 only. Set at 110ꢀ–132ꢀ of
nominal
Overcurrent Protection
Temperature Coefficient
All outputs protected for short circuit conditions
All outputs 0.04ꢀ per ꢁC
Holdup Time
PFD Signal
Typically 12mS at 110VAC
Power fail detection signal. TTL logic low upon
loss of input power.
340