Medline 200 Series
1 5 0 - 2 0 0 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L
Features
• Low safety ground leakage current
• Wide input range 90 to 260Vac
• Input surge current protection
• Overvoltage protection
• Overcurrent protection
• Power fail detection (PFD) signal
• Optional cover and fan assembly
• 100% burn-in at full rated load
• Remote inhibit - TTL high disables output
• Short curcuit protection
PFD signal
TTL logic high for normal operation and TTL logic low
Specifications
upon loss of input power.This signal appears at least
1ms prior to +5V output dropping 5% below its
nominal value.This signal also provides a minimum
delay of 100ms after +5V is within regulation
INPUT
Voltage range
Frequency
90-260VAC
47-63Hz.
Efficiency
70% min on all models
20ms min at 110VAC
0.5% maꢀ at full load
Input Current
3.2A (rms) at 115VAC
1.6A (rms) at 230VAC
Hold up time
Leakage current
90µA maꢀ at 115VAC, 60Hz
150A maꢀ at 230VAC, 50Hz
Line regulation
ENVIRONMENTAL
Inrush Current
Isolation
20A @ 115V or 40A at 230V at 25°Cand cold start
Input – Output: 4000VAC
Input – Ground: 1500VAC
Output – Ground: 500VAC
Operating Temperature
Storage temperature
Humidity
0°C to 70°C
–40°C to +85°C
5% to 95% non-condensing
OUTPUT
Derating
Derate from 100% at +50°C linearly to 50% at
+70°C.
Power
200W
Voltage
See table
See table
2% p-p maꢀ
Cooling
25CFM forced air, provided on “C” version, to be
provided for “B” version by user
Current
GENERAL
Ripple & Noise
Overvoltage protection
Switching frequency
MTBF
88-112kHz.
Provided on output #1 only, set at 112-132% of it
nominal output voltage
350,000 hours at full load 25°C ambient ,
calculated per MIL-HDBK-217F
Overcurrent Protection
Temperature coefficient
Transient response
The output protected to short circuit conditions
0.04%°Call outputs
STANDARDS AND APPROVALS
Maꢀ eꢀcursion of 4% or better on all models,
recovering to 1% of final value within 500us after
a 25% step load change
Safety standards
C-Tick
IEC60601-1 TUV EN60601-1, EN60601-1
AS/NZS CISPR11 Group 1, Class A
EMC standards
EN55011 Class B, FCC Class B,VCII, EN61000-3-2, -3,
EN60001-4-2, -3, -4, -5, -6, -8, -11
Technical Illustration
MECHANICAL
Dimensions
Weight
See drawing
0.82kg for “B” version
0.96kg for version “C” version
Notes:
1. Dimensions shown in inches (mm)
2. Tolerance 0.02 (0.5) maximum
3. Input connector P1 is Dinkle DT-35-B01W-09 screws are M3, nickle plated
4. Connector P3 mates with Molex housing 22-01-1043 and Molex 40445 series
crimp terminal
5. Connector P4/P5 mates with Molex housing 22-01-1023 and Molex 40445
series crimp terminal
6. P4 is for DC fan rated 24V/0.2A (models PM201/27), or 5V/0.38A (models
PM201-40-3), and 12V/0.2A (other models, Pin1 +V and Pin2 COM
7. Weight:
0.82 Kgs (1.83 Lbs) approx for “B” version
0.96 Kgs (2.14 Lbs) approx for “C” version
345
Your dependable power partner – www.powerbox.com.au