1
5 ⁄ Digit Multimeters
2
Specifications – NI 4060
Specifications are guaranteed between 15 and 35 °C unless otherwise noted.
AC Current (Accuracy % of reading mA)
24 Hour
(25 °C 1 °C)
90 Day
1 Year
Temperature Coefficient
DC Voltage (Accuracy % of reading µV)
Range
(25 °C 10 °C) (25 °C 10 °C) (% of reading/°C mA/°C)
24 Hour
90 Day
1 Year
Temperature Coefficient
20.0000 mA 0.30% + 100 µA 0.32% + 170 µA 0.32% + 170 µA 0.022%/°C + 12 µA/°C
200.000 mA 0.18% + 0.22 mA 0.20% + 1.2 mA 0.20% + 1.2 mA 0.009%/°C + 0.110 mA/°C
10.0000 A* 0.3% + 22 mA 0.32% + 120 mA 0.32% + 120 mA 0.026%/°C + 11 mA/°C
Range
(25 °C 1 °C)
(25 °C 10 °C) (25 °C 10 °C) (% of reading/°C + µV/°C)
250.000 V
25.0000 V
2.00000 V
0.0032% + 1.25 mV 0.021% + 1.25 mV 0.024% + 1.25 mV 0.0017%/°C + 480 µV/°C
0.0032% + 1 mV
0.0029% + 10 µV
0.021% + 1 mV
0.024% + 1 mV
0.0017%/°C + 480 µV/°C
*Requires 10 A shunt, CSM-10A
Accuracy numbers are for 5 ⁄
errors, temperature variation, linearity, and noise.
0.014% + 10 µV 0.017% + 10 µV 0.0009%/°C + 5 µV/°C
1
2
digits and include the effects of full and zero-scale
200.000 mV 0.0029% + 6 µV
20.0000 mV 0.0029% + 6 µV
0.014% + 6 µV
0.014% + 6 µV
0.017% + 6 µV
0.017% + 6 µV
0.0009%/°C + 1 µV/°C
0.0009%/°C + 1 µV/°C
1
Input Characteristics
Accuracy numbers are for 5 ⁄ digits with autozero on and include the effects of
2
Input protection ......................................... Fuse F1 500 mA/250 V fast-fusing
Shunt resistor ............................................ 1 Ω
Burden voltage........................................... < 400 mV at 200 mA AC
full and zero-scale errors, temperature variation, linearity, and noise. Overrange
capabilities to 300 V.
Noise Rejection
Resistance (Accuracy % of reading Ω)
NMRR ........................................................ 80 dB (10 Hz filter setting, 50/60 Hz
powerline frequency 1ꢀ)
24 Hour
90 Day
1 Year
Temperature Coefficient
Range
(25 °C 1 °C)
(25 °C 10 °C) (25 °C 10 °C) (% of reading/°C + Ω/°C)
DC ECMRR................................................ 140 dB (with a 1 kΩ imbalance in HI lead)
AC ECMR (50/60 Hz) ................................. 150 dB (with a 1 kΩ imbalance in HI lead)
Extended ohm
(> 2 MΩ)
0.1% + 6 kΩ
0.012% + 9 Ω
0.012% + 5 Ω
0.1% + 60 kΩ
0.077% + 27 Ω 0.080% + 27 Ω
0.077% + 22 Ω 0.080% + 22 Ω
0.1% + 60 kΩ
0.0072%/°C + 6 kΩ/°C
0.0072%/°C + 2 Ω/°C
0.0072%/°C + 2 Ω/°C
2.00000 MΩ*
200.000 kΩ
20.0000 kΩ
2.00000 kΩ
200.000 Ω
Input Characteristics
Input bias current....................................... 1 nA maximum
Input resistance ......................................... >1 GΩ (2 V, 200 mV, 20 mV ranges);
1 MΩ (250 V, 25 V)
0.006% + 0.09 Ω 0.024% + 0.3 Ω 0.027% + 0.3 Ω 0.0020%/°C + 0.02 Ω/°C
0.006% + 0.05 Ω 0.024% + 0.2 Ω 0.027% + 0.2 Ω 0.0020%/°C + 0.02 Ω/°C
0.006% + 0.05 Ω 0.024% + 0.2 Ω 0.027% + 0.2 Ω 0.0020%/°C + 0.02 Ω/°C
DC Current (Accuracy % of reading µA)
*With autozero on, or while scanning, and when large resistance with capacitive
loads is measured, additional delay time is required.
24 Hour
90 Day
1 Year
Temperature Coefficient
1
Accuracy numbers are for 4-wire resistance mode, 5 ⁄
2
digits with autozero on and
Range
(25 °C 1 °C)
0.015% + 10 µA
0.015% + 10 µA
0.02% + 1 mA
(25 °C 10 °C) (25 °C 10 °C) (% of reading/°C + µA/°C)
include the effects of full and zero-scale errors, temperature variation, linearity, and noise.
Measurement modes
Ohms.................................................... 2 or 4-wire
Extended ohms .................................... 2-wire only
Maximum lead resistance.......................... 10 Ω (200 Ω range), 1 kΩ
(all other ranges)
20.0000 mA
200.000 mA
10.0000 A*
0.039% + 10 µA 0.042% + 10 µA
0.039% + 10 µA 0.042% + 10 µA
0.035% + 2 mA 0.035% + 2 mA
0.0035%/°C + 1 µA/°C
0.0035%/°C + 1 µA/°C
0.007%/°C + 0.1 mA/°C
*Requires 10 A shunt, CSM-10A
1
Accuracy numbers are for 5 ⁄
2
digits with autozero on and include the effects of
full and zero-scale errors, temperature variation, linearity, and noise.
Additional error for 2-wire resistance ........ 0.6 Ω
Input Characteristics
Input protection ......................................... Fuse F1 500 mA/250 V fast-fusing
Shunt resistor ............................................ 1 Ω
Diode Testing (Accuracy % of reading µV)
24 Hour
(25 °C 1 °C)
0.006% + 7 µV
90 Day
(25 °C 10 °C) (25 °C 10 °C) (% of reading/°C + µV/°C)
0.024% + 22 µV 0.027% + 22 µV 0.0020%/°C + 2 µV/°C
1 Year
Temperature Coefficient
Burden voltage........................................... <400 mV at 200 mA DC
Range
2 V
AC Voltage (Accuracy % of reading mV)
1
Accuracy numbers are for 5 ⁄ digits with autozero on and include the effects of
2
24 Hour
90 Day
(25 °C 10° C) (25 °C 10 °C) (% of reading/°C + mV/°C)
0.62% + 680 mV 0.62% + 680 mV 0.007%/°C + 20 mV/°C
0.18% + 210 mV 0.18% + 210 mV 0.007%/°C + 20 mV/°C
0.30% + 21 mV 0.30% + 21 mV 0.019%/°C + 2 mV/°C
1 Year
Temperature Coefficient
full and zero-scale errors, temperature variation, linearity, and noise.
Test current................................................ 100 µA
Range
(25 °C 1 °C)
0.6% + 250 mV
0.16% + 30 mV
0.28% + 3 mV
250.000 V
25.0000 V
2.00000 V
200.000 mV
20.0000 mV
General Specifications
Settling time .............................................. Affected by source impedance and
input signal changes.
0.16% + 0.22 mV 0.18% + 1.20 mV 0.18% + 1.20 mV 0.007%/°C + 0.110 mV/°C
0.28% + 100 µV 0.30% + 170 µV 0.30% + 170 µV 0.019%/°C + 12 µV/°C
Warm-up time............................................ 30 minutes for measurements
accurate within specifications.
Altitude....................................................... Up to 2,000 m; at higher altitudes the
installation category must be derated.
Working voltage......................................... 300 V maximum between either input
terminal and earth ground
1
Accuracy numbers are for 5 ⁄ digits and include the effects of full and zero-scale
2
errors, temperature variation, linearity, and noise. Applies for sine waves ≥ 10ꢀ of
input range. Accuracy may be affected by source impedance, cable capacitances,
dielectric absorption, or slew rate. Overrange capabilities to 300 V.
Noise Rejection
Power Requirement
+5 VDC ...................................................... 250 mA in operational mode
AC CMRR (50/60 Hz) ................................ >80 dB (with a 1 kΩ imbalance in HI lead)
Input Characteristics
Input resistance ......................................... 1 MΩ
Bandwidth.................................................. 20 Hz - 25 kHz
Physical
Dimensions
PCI........................................................ 10.8 by 17.5 cm (4.25 by 6.9 in.)
PXI ........................................................ 10 by 16 cm (3.9 by 6.33 in.)
Additional AC Errors
Frequency dependent errors
Environment
Operating temperature ..............................
0 to 55 °C
Input Frequency Additional Error (% of reading)
Storage temperature.................................. -20 to 70 °C
Relative humidity ....................................... 10 to 90ꢀ, noncondensing
20 Hz - 50 Hz
50 Hz - 100 Hz
100 Hz - 20 kHz
20 Hz - 25 kHz
2.5%
0%
1%
Certifications and Compliances
2.5%
CE Mark Compliance
National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • info@ni.com • ni.com 465