MUU61 SERIES
63W U-bracket Switching Power Supplies For Medical Equipment.
Description:
The MUU61 series of compact, open frame constructed, AC/DC
switching mode power supplies provide 63 Watts of continuous
output power . They are ideally suited for use in hospital instrument
and many other applications. All models meet FCC Part-18 class B
and CISPR-11 EN55011 class B emission Limits and are designed
to comply with UL/c-UL(UL 60601-1) ,TUV/T-mark(EN 60601-1)
and new CE requirements. All units are 100% burned in and tested.
Features:
Wide Input Voltage 90 to 264 VAC,47 to 63 Hz
Internal EMI filter
Single to Quad Output
Input connector mates with Molex housing 09-50-3031 and
Molex 2478 series crimp terminal
Output connector mates with Molex housing 09-50-3081
and Molex 2478 series crimp terminal
Output Voltage Available From 3 VDC Thru 40 VDC
Input Surge Current, Over Voltage and Over Load protection
Power Fail Detect (Optional)
Class I Insulation
Safety Approvals :
Two Years Warranty
Electrical Characteristics:
Sym.
Parameter
Test Conditions
Typ.
Min.
Max. Unit
Vin
Fin
Wo
Input Voltage
Operating Voltage
264
63
VAC
Hz
W
90
Input Frequency
47
Output Power Range
63
Vin=90 to 264 VAC
0
Vo
Output Voltage Range
Output Current Range
Input Current (Low Line)
Input Current (High Line)
V
See rating Chart
Io
See rating Chart
A
Iil
Iih
Io=Full load, Vin=115VAC
Io=Full load, Vin=230VAC
1.6
0.8
A
A
Io=Full load, 25 C,Cool start,
Vin=115VAC
Low Line Inrush Current
High Line Inrush Current
Irl
Irh
A
15
21
18
25
Io=Full load, 25 C,Cool start,
Vin=230VAC
A
%
Efficiency
Eff
Io=Full load, Vin=230VAC
Io=Full load
85
1
70
80
0.5
3
Line Regulation
REG-i
%
%
REG-o Load Regulation
Vin=230VAC
6
Over Voltage Protection
OVP
OCP
Ttr
112
110
%
132
150
4
Over Current Protection
Transient Response
%
Io=Full load to Half Load, Vin=100VAC
Io=Full load, Vin=110VAC
mS
mS
S
Thold Hold-Up Time
16
Ts
Start Up Time
Io=Full load, Vin=100VAC
1
1
2
2
0.3
Vp-p
Ripple & Noise (Peak to Peak) Full load, Vin=90VAC
Safety Ground Leakage Current Io=Full load, Vin=240VAC
%
mA
Ilk
Tc
0.1 0.3
All output
Temperature Coefficient
-0.04
0.04 %/ C
Environmental :
Sym.
Typ.
Test Conditions
Parameter
Unit
Min.
0
Max.
70
Toper Operating Temperature
C
C
%
Tstg
Hr
-40
5
85
Storage Temperature
Relative Humidity
Derate linearly from 100% load at 50 C to 50% load at 70 C
95
Pd
Safety Specifications:
Sym.
Parameter
Min. Typ. Max. Unit
Test Conditions
Dielectric Withstanding Voltage
Vps
VDC
Primary to secondary
5600
for Primary to secondary
Dielectric Withstanding Voltage
for Primary to Ground
Primary to ground
Vpg
Ri
2800
VDC
Isolation Resistance
Test Voltage=2100VDC
Vin=220VAC
M
50
B
CISPR EMI requirements for CISPR-11
CLASS
CLASS
EMIrequirementsforFCCPART-18
FCC
Vin=110VAC
B