AC-DC ITE & Medical Switching Power Supply
MQF240U SERIES
240 Watts
KEY FEATURES
Standard
A Type
U Bracket Medical Switching Power Supply
Cooling by Free Air Convection
160 Watts and 240 Watt with 10CFM Forced Air
4000VAC Input to Output 2MOPP Insulation
High Efficiency up to 94%
With P.F.C. Function >0.9
<0.5W No Load Input Power
Built-in 12V / 0.5A Fan Supply
EMI for Both Class I (with PE) and Class II (without PE) Configuration
Suitable for BF Application with Appropriate System Consideration
UL / IEC / EN 60601 3.1 Edition & UL / IEC / EN 60950 AM2 Safety Approvals
3-Year Product Warranty
(except for 15S)
ELECTRICAL SPECIFICATIONS
All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Model No.
MQF240U-12S
240 W
MQF240U-15S
MQF240U-24S
MQF240U-48S
Max Output Wattage (with 10CFM FAN) (W)
160 W
Max Output Wattage (Free air Convection) (W)
(Note 4)
Voltage
90-264 VAC
47-63 Hz
Frequency (Hz)
Current (Full load)
< 3.0 A max. (115 VAC) / < 1.5 A max. (230 VAC)
< 45 A max. (115 VAC) / < 90 A max. (230 VAC)
< 0.1mA / 264 VAC (Touch Current)
PF>0.9 at Full Load
Input
Inrush Current (<2ms)
Leakage Current
Power Factor
No Load
< 0.5W (115 / 230 VAC)
Voltage (V.DC.)
12V
15V
24V
48V
Voltage Adj Range (V.DC.)
Voltage Accuracy
±4% Output Voltage
±2%
Current (with 10CFM FAN) (A) (max.)
Current (Free air Convection) (A) (max.)
Line Regulation
20
16
10
5
13.3
10.667
6.66
3.33
±1%
Output
Load Regulation (0-100%)
Minimum Load
±1%
0%
Maximum Capacitive Load
8000µF
1% Vout
92.5%
10 ms min.
2000µF
92.5%
3000µF
93%
470µF
94%
(Note 1)
(Note 6)
(Note 2)
Ripple & Noise (max.)
Efficiency (at 230VAC)
Hold-up Time (at 115 VAC)
Over Power Protection
Over Voltage Protection
Overt Temperature Protection
Auto recovery, Hiccup mode
Auto recovery
Protection
Isolation
Auto recovery
Protection level 1 (nominal):Continuous, Auto recovery
Protection level 2 (instantaneous high current):Latch
4000VAC or 5656VDC
Short Circuit Protection
(Note 5)
(Note 5)
(Note 5)
Input-Output
Input-PE
2000VAC or 2828VDC
Output-PE
1500VAC or 2121VDC
Operating Temperature
Storage Temperature
Temperature Coefficient
Altitude During Operation
Humidity
-30°C…+70°C (with derating)
-30°C…+85°C
±0.05%/°C
5000m
Environment
20~90% RH
Atmospheric Pressure
MTBF
56 kPa to 106 kPa
>250,000 h @ 25°C (MIL-HDBK-217F, Notice 1)
10~500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes.
Vibration
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
-1-