5秒后页面跳转
MQF120E-12S PDF预览

MQF120E-12S

更新时间: 2024-09-19 01:04:31
品牌 Logo 应用领域
ARCH /
页数 文件大小 规格书
3页 204K
描述
AC-DC Medical Switching Power Supply

MQF120E-12S 数据手册

 浏览型号MQF120E-12S的Datasheet PDF文件第2页浏览型号MQF120E-12S的Datasheet PDF文件第3页 
VER : A_2  
update : 2018.06.06  
AC-DC Medical Switching Power Supply  
MQF120E SERIES  
120 Watts  
KEY FEATURES  
Enclosed Medical Switching Power Supply  
Cooling by Free Air Convection  
85 Watts and 120 Watt with 10CFM Forced Air  
4000VAC Input to Output 2MOPP Insulation  
High Efficiency up to 91%  
With P.F.C. Function >0.9  
<0.3W No Load Input Power  
EMI for Both Class I (with PE) and Class II (without PE) Configuration  
Suitable for BF Application with Appropriate System Consideration  
UL / IEC / EN 60601 3.1 Edition Safety Approvals  
3-Year Product Warranty  
(In Progress)  
ELECTRICAL SPECIFICATIONS  
All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated.  
Model No.  
MQF120E-12S  
120 W  
MQF120E-24S  
MQF120E-48S  
Max Output Wattage (with 10CFM FAN) (W)  
85 W  
Max Output Wattage (Free air Convection) (W)  
(Note 3)  
Voltage  
90-264 VAC  
47-63 Hz  
Frequency (Hz)  
Current (Full load)  
< 2.0 A max. (115 VAC) / < 1.0 A max. (230 VAC)  
< 45 A max. (115 VAC) / < 90 A max. (230 VAC)  
< 0.1mA / 264 VAC (Touch Current)  
PF>0.9 at Full Load  
Input  
Inrush Current (<2ms)  
Leakage Current  
Power Factor  
No Load  
< 0.3W (115 / 230 VAC)  
Voltage (V.DC.)  
12V  
24V  
48V  
Voltage Adj Range (V.DC.)  
Voltage Accuracy  
±4% Output Voltage  
±2%  
Current (with 10CFM FAN) (A) max  
Current (Free air Convection) (A) max  
Line Regulation  
10  
5
2.5  
7.083  
3.542  
1.771  
±1%  
Output  
Load Regulation (10-100%)  
Minimum Load  
±1%  
0%  
Maximum Capacitive Load  
3000µF  
1500µF  
1% Vout  
90%  
500µF  
91%  
(Note 1)  
Ripple & Noise (max.)  
Efficiency (at 230VAC)  
Hold-up Time (at 115 VAC)  
Over Power Protection  
Over Voltage Protection  
Overt Temperature Protection  
Short Circuit Protection  
Input-Output  
160mV  
90%  
(Note 2)  
10 ms min.  
Auto recovery, Hiccup mode  
Latch off  
Protection  
Isolation  
Latch off  
Auto recovery, Hiccup mode  
4000VAC or 5656VDC  
2000VAC or 2828VDC  
1500VAC or 2121VDC  
-30°C…+70°C (with derating)  
-30°C…+85°C  
±0.05%/°C  
Input-FG  
Output-FG  
Operating Temperature  
Storage Temperature  
Temperature Coefficient  
Altitude During Operation  
Humidity  
5000m  
Environment  
20~90% RH  
54 kPa to 106 kPa  
Atmospheric Pressure  
MTBF  
>250,000 h @ 25°C (MIL-HDBK-217F, Notice 1)  
10~500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes.  
Vibration  
http://www.archcorp.com.tw  
TEL: +886-2-26989508 FAX: +886-2-26981319  
-1-  

与MQF120E-12S相关器件

型号 品牌 获取价格 描述 数据表
MQF120E-24S ARCH

获取价格

AC-DC Medical Switching Power Supply
MQF120E-48S ARCH

获取价格

AC-DC Medical Switching Power Supply
MQF120O ARCH

获取价格

Open Frame Medical Switching Power Supply
MQF120O_18 ARCH

获取价格

AC-DC Medical Switching Power Supply
MQF120O-12S ARCH

获取价格

Open Frame Medical Switching Power Supply
MQF120O-24S ARCH

获取价格

Open Frame Medical Switching Power Supply
MQF120O-48S ARCH

获取价格

Open Frame Medical Switching Power Supply
MQF120U ARCH

获取价格

AC-DC Medical Switching Power Supply
MQF120U_18 ARCH

获取价格

AC-DC Medical Switching Power Supply
MQF120U-12S ARCH

获取价格

AC-DC Medical Switching Power Supply