DTMPU63 Series
63 Watt
Single Output
AC/DC Medical Desktop Power Supply
FEATURES
● Splash Proof
● 3 Year Warranty
● Class I Insulation
● IEC 320-C14 Input Inlet
● Optional Output Connectors
● CEC and Energy Star Compliant
● Output Voltage Protection (Crowbar Design)
● Wide Input Voltage 90 to 264VAC, 47 to 63Hz
● Output Voltages Available from 5VDC to 36VDC
● Input Surge Current, Over Voltage, and Over Load Protection
SPECIFICATIONS: DTMPU63 Series
All specifications are based on 25oC, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Nom
Max
Unit
INPUT (Vin)
Input Voltage
90
47
264
63
1.62
0.72
15
30
0.3
2
VAC
Hz
A
A
A
A
mA
s
Input Frequency
Input Current (Low Line)
Input Current (High Line)
Inrush Current (Low Line)
Inrush Current (High Line)
Safety Ground Leakage Current
Start-Up Time
Io = Full Load, Vin = 115VAC
Io = Full Load, Vin = 230VAC
Io = Full Load, 25°C, Cool Start, Vin = 115VAC
Io = Full Load, 25°C, Cool Start, Vin = 230VAC
Io = Full Load, Vin = 240VAC
12
26
0.1
1
Io = Full Load, Vin = 100VAC
0.3
OUTPUT (Vo)
Output Voltage Range
Load Regulation
See Rating Chart
VDC
%
%
Vin = 230VAC
Io = Full Load
3
0.5
7
1
Line Regulation
Output Power
Vin = 90 to 264VAC
0
63
W
A
%
ms
ms
%/°C
Output Current Range
Ripple & Noise (peak to peak)
Transient Response
Hold-Up Time
Temperature Coefficient
PROTECTION
See Rating Chart
0.5
Full Load, Vin = 90VAC
1
4
Io = Full Load to Half Load, Vin = 100VAC
Io = Full Load, Vin = 110VAC
All Outputs
16
-0.04
+0.04
Over Voltage Protection
Over Current Protection
GENERAL
112
110
132
150
%
%
Efficiency
Io = Full Load, Vin = 230VAC
Primary to Secondary
77
85
88
%
Dielectric Withstanding Voltage
For Primary to Secondary
Dielectric Withstanding Voltage
For Primary to Ground
Isolation Resistance
5600
VDC
Primary to Ground
2800
50
VDC
Test Voltage = 2100VDC
MΩ
ENVIRONMENTAL
Operating Temperature
Storage Temperature
Relative Humidity
MTBF
Derate linearly from 100% Load at 50°C to 50% load at 70°C
0
-40
5
50
+70
+85
95
°C
°C
%
100,000
hours
PHYSICAL
Weight
Approximately 510 ~ 560
5.75(L) x 2.99(W) x 1.69(H)
3
grams
inches
years
Dimensions
Warranty
SAFETY
EMI Requirements for CISPR-11
EMI Requirements for FCC PART-18
Vin = 220VAC
Vin = 110VAC
B
B
Class
Class
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833