1S7B_6UP series
1W - Single/Dual Output DC-DC Converter - Fixed Input - Isolated & Unregulated
DC-DC Converter
1 Watt
The 1S7B_6UP series are specially designed for applications where an isolated
voltage is required in a distributed power supply system. They are suitable for:
pure digital circuits, low frequency analog circuits, relay-driven circuits and data
switching circuits.
High efficiency up to 80%
4200VAC/6000VDC isolation
SIP package
Operating temperature:
-40°C up to +85°C
RoHS compliance
Reinforced insulation
The patient leakage current:
Max 2μA
Short circuit protection
EN60601-1, ANSI/AAMI
ES60601-1 approved
(1 x MOPP/2 x MOOP)
Meets EN62368
These products apply to:
1) Where the voltage of the input power supply is fixed (voltage
variation ≤ 10%)
International standard pinout
2) Where isolation is necessary between input and output
(isolation voltage ≤4200VAC or 6000VDC)
3) Where do not has high requirement of line regulation and the ripple & noise
of the output voltage; such as: Medical collection and isolation, High voltage
collection circuit, IGBT-driven circuits, etc.
UL-60601-1 (E486991)
Common specifications
Short circuit protection:
Operation temperature range:
Storage temperature range:
Pin Soldering Resistance Temp
Storage humidity range:
MTBF (MIL-HDBK-217F@25°C):
Case Temperature Rise (Ta=25°C)
Case material:
Output specifications
Continuous, self-recovery
-40°C – +105°C
Item
Test condition
Min Typ
Max Units
Voltage accuracy See output regulation curve (Fig. 1)
-55°C – +125°C
Line regulation
Load regulation
For Vin change of 1%
1.2
%
300°C MAX, 1.5mm from case for 10 sec
< 95% (Non-condensing)
>3,500,000 hours
10% to 100% load
• 3.3VDC output
• 5VDC output
15
12
8
%
%
%
%
%
%
• 7.2/9VDC output
• 12VDC output
• 15VDC output
• 24VDC output
7
25°C
6
5
Black plastic; flame-retardant and
heat-resistant (UL94-V0)
Temperature
coefficient
full load
0.03
%/°C
Weight:
4.2g
Ripple & Noise*
20MHz Bandwidth
• 3.3/5/7.2/9/12VDC output
• 15/24VDC output
Dimensions
Cooling:
19.50 x 9.80 x 12.50 mm
Free air convection
100
150
150
120
mVp-p
mVp-p
Switching
frequency
Full load, nominal input
50
KHz
Input specifications
* Note:* The “parallel cable” method is used for Ripple and Noise test, please
refer to DC-DC Converter Application Notes for specific information.
Item
Test condition
Min Typ
Max
Units
Input current
(no-load/full load)
• 5VDC input
3.3VDC output
5/7.2VDC output
9/12/15VDC output
24VDC output
• 12VDC input
5/7.2VDC output
9/15VDC output
12VDC output
• 15VDC input
5VDC output
278/30 295/-
256/30 271/-
253/30 267/-
250/30 264/-
mA
mA
mA
mA
Isolation specifications
Item
Test condition
Min
Typ
10
Max
Units
VDC
Isolation voltage
Tested for 1 minute and 6000
1mA max
104/20 110/-
102/20 107/-
103/20 109/-
mA
mA
mA
Isolation resistance
Test at 500VDC
1000
MΩ
pF
Isolation capacitance Input/Output,
100KHz/0.1V
8
2
82/15
81/15
88/-
87/-
mA
mA
15VDC output
• 24VDC input
3.3VDC output
5/12/15VDC output
Patient leakage
current*:
250VAC, 50/60Hz
µA
58/10
52/10
62/-
55/-
mA
mA
Reflected Ripple
Current*
• 5VDC input
15
5
mA
mA
• 12/15/24VDC input
Example:
1S7B_0505D6UP
Input surge
voltage
(1sec. max.)
• 5VDC input
• 12VDC input
• 15VDC input
• 24VDC input
-0.7
-0.7
-0.7
-0.7
9
VDC
VDC
VDC
VDC
1 = 1Watt; S7 = SIP7; B = Pinning; 05 = 5Vin; 05 = 5Vout;
D = Dual Output; 6 = 6kVDC isolation; U = Unregulated Output;
P = Short circuit protection
18
21
30
Input filter
Hot plug
Capacitor
Unavailable
EMC specifications
Note: * Refer to DC-DC Converter Application Notes for detailed description of
reflected ripple current test method.
EMI
EMI
EMS
CE
CISPR32/EN55032 CLASS B
(see EMC recommended circuit)
RE
CISPR32/EN55032 CLASS B
(see EMC for recommended circuit)
ESD
IEC/EN61000-4-2 Contact 8kV perf. Criteria B
Page 1 of 4
1S7B_6UP – Rev. 2020-2.1
Specifications subject to change without notice.