TRUMPower
110 WATT MEDICAL POWER ADAPTERS
TMP110 SERIES
DESCRIPTION
(MDD)
The TMP110 series of AC/DC switching power
adapters are for 110 watts of continuous
output power. They are enclosed in a 94V-1
rated polyphenylene-oxide case with an IEC
320 inlet to mate with interchangeable cord for
world-wide use. An on/off switch can be added
at request. All models meet CISPR 11 and FCC
class B emission limits, and are designed for
medical applications.
FEATURES
‧
‧
‧
‧
‧
‧
‧
‧
‧
Dual, triple or quadruple outputs
Optional output connectors
Optional on/off switch
Safety Standard Approvals:
Optional Power Fail Detect (PFD) signal
100% burn-in
UL 2601-1
CSA C22.2 No. 601.1
Wide input range 85 to 264 VAC
Input surge current protection
Overvoltage protection
TÜV EN60601-1
Overcurrent protection
INPUT SPECIFICATIONS
ENVIRONMENTAL SPECIFICATIONS
Input voltage:
85 to 264 VAC
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
0ºC to +70ºC
Input frequency: 47 to 63 Hz
-40ºC to +85ºC
Input current:
3.20A (rms) for 115 VAC
1.80A (rms) for 230VAC
5% to 95% non-condensing
Derate from 100% at +50ºC
linearly to 50% at +70ºC
Leakage current: 90µA max. @ 115VAC, 60Hz
150µA max. @ 230VAC, 50Hz
OUTPUT SPECIFICATIONS
Output voltage/current: See rating chart
GENERAL SPECIFICATIONS
Switching frequency: 20KHz to 250KHz, varies with load and line
Efficiency:
70% minimum on single output models with
Vo ≥12V, 65% minimum on the others
12 msec minimum at 110VAC
Total output power:
Ripple and noise:
110 watts maximum
1% peak to peak maximum
Hold-up time:
Overvoltage protection: Provided on output #1 only;
set at 112-132% of its nominal
Line regulation:
±0.5% maximum at full load
Inrush current:
15 amps @ 115VAC, or 30 amps
@230VAC, at 25ºC cold start
4000VAC from input to output 1500VAC
from input to ground
500VAC from output to ground
550,000 hours minimum at full load at 25ºC
ambient, calculated per MIL-HDBK-217F,
excluding DC fan.
output voltage
Withstand voltage:
Overcurrent protection: All outputs protected to short
circuit conditions
MTBF:
Temperature coefficient: All outputs ±0.04%/ºC
maximum
EMC Performance (EN60601-1-2: 2001)
Transient response:
Maximum excursion of 4% or
better on all models,
recovering to 1% of final
value within 500us after a
25% step load change
EN55011:
FCC:
VCCI:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Harmonic distortion, Class A
Line flicker
ESD, ±8KV air and ±6KV contact
PFD signal (optional):
TTL logic high for normal
operation and TTL logic low
upon loss of input power. This
signal appears at least 1ms
prior to +5V output dropping
5% below its nominal value.
This signal also provides a
minimum delay of 100ms after
+5V is within regulation.
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
EN61000-4-11:
Radiated immunity, 3V/m for 80~2500MHz
Fast transient/burst, ±2KV
Surge, ±1KV diff, ±2KV com.
Conducted immunity, 3Vrms
Magnetic field immunity, 3A/m
Voltage dips, 30% reduction for 500ms, 60%
reduction for 100ms and >95% reduction for
10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com