25-28 WATT MEDICAL POWER SUPPLIES

TM25-S28 数据手册

 浏览型号TM25-S28的Datasheet PDF文件第2页浏览型号TM25-S28的Datasheet PDF文件第3页 
TRUMPower  
The TM25 series of compact, open PCB constructed,  
AC-DC switching power supplies are capable of  
delivering 25 to 28 watts of continuous output  
power. They operate at 85 to 264VAC input voltage  
without the need of voltage selection. They are  
ideally suited for use in medical equipments, safety  
systems and monitoring equipments.  
25-28 WATT MEDICAL POWER SUPPLIES  
DESCRIPTION  
(LVD)  
TM25 SERIES  
RoHS  
FEATURES  
Low safety ground leakage current  
100% burn-in  
85 to 264VAC wide input range  
Input surge current protection  
Overvoltage protection  
Safety Standard Approvals:  
UL 2601-1  
CSA C22.2 No. 601.1  
Overcurrent protection  
Open PCB construction  
Compliant with RoHS requirements  
TÜV EN60601-1  
INPUT SPECIFICATIONS  
Input voltage:  
Input frequency:  
Input current:  
85 to 264VAC  
GENERAL SPECIFICATIONS  
47 to 63 Hz  
0.7Armsfor 115VAC  
0.4Armsfor 230VAC  
Switching frequency42KHz ±5KHz  
Efficiency:  
70% minimum on single output  
model with Vo 12V, 68% minimum  
on the others  
Earth leakage current90µA max. @ 115VAC, 60Hz  
150µA max. @ 230VAC, 50Hz  
Hold-up time:  
Line regulation:  
Inrush current:  
10 msec minimum at 110VAC  
±0.5% maximum at full load  
OUTPUT SPECIFICATIONS  
10 amps @ 115VAC or 25 amps  
@ 230VAC at 25ºC cold start  
Output voltage/currentSee Rating Chart  
Total output power:  
Ripple and noise:  
See Rating Chart  
Withstand voltage:  
MTBF:  
4000VAC from input to output  
1500VAC from input to ground  
500VAC from output to ground  
1% peak to peak max.  
Overvoltage protectionProvided on output #1 only;  
set at 112-132% of its  
500,000 hours minimum at full load  
at 25ºC ambient, calculated per  
MIL- HDBK- 217F  
nominal output voltage  
Overcurrent protectionAll outputs protected to short  
circuit conditions  
EMC Performance (EN60601-1-2:2001)  
Temperature coefficientAll outputs ±0.04%/ºC  
EN55011:  
FCC:  
Class B conducted, Class B radiated  
Class B conducted, Class B radiated  
maximum  
Transient response:  
Maximum excursion of 4%  
or better on all models,  
recovering to 1% of final  
value within 500µs after a  
25% step load change  
VCCI:  
Class B conducted, Class B radiated  
EN61000-3-2:  
EN61000-3-3:  
EN61000-4-2:  
EN61000-4-3:  
EN61000-4-4:  
EN61000-4-5:  
EN61000-4-6:  
EN61000-4-8:  
Harmonic distortion, Class A and D  
Line flicker  
ESD, ±8KV air and ±6KV contact  
Radiated immunity, 3V/m for 80~2500 MHz  
Fast transient/burst, ±2KV  
ENVIRONMENTAL SPECIFICATIONS  
Surge, ±1KV diff, ±2KV com.  
Conducted immunity, 3Vrms  
Magnetic field immunity, 3A/m  
Operating temperature: 0ºC to +70ºC  
Storage temperature-40ºC to +85ºC  
Relative humidity:  
Derating:  
5% to 95% non-condensing  
EN61000-4-11: Voltage dips, 30% reduction for 500ms,  
60% reduction for 100ms and >95%  
reduction for 10ms  
Derate from 100% at +50ºC  
linearly to 50% at +70ºC  
Tumbler Technologies + TRUMPower  
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA  
Phone: 408-988-6616 Fax: 408-988-6622 email: sales@trumpower.com Website: www.TRUMPower.com  

与TM25-S28相关器件

型号 品牌 获取价格 描述 数据表
TM25-S5.1 TRUMPOWER

获取价格

25-28 WATT MEDICAL POWER SUPPLIES
TM25-T31 TRUMPOWER

获取价格

25-28 WATT MEDICAL POWER SUPPLIES
TM25-T32 TRUMPOWER

获取价格

25-28 WATT MEDICAL POWER SUPPLIES
TM25-T39 TRUMPOWER

获取价格

25-28 WATT MEDICAL POWER SUPPLIES
TM25T3A-H MITSUBISHI

获取价格

MEDIUM POWER GENERAL USE INSULATED TYPE
TM25T3A-M MITSUBISHI

获取价格

MEDIUM POWER GENERAL USE INSULATED TYPE
TM266DCA PULSE

获取价格

MILITARY/AEROSPACE PRODUCTS
TM266DCA1 PULSE

获取价格

MILITARY/AEROSPACE PRODUCTS
TM266DCA2 PULSE

获取价格

MILITARY/AEROSPACE PRODUCTS
TM266DCA5 PULSE

获取价格

MILITARY/AEROSPACE PRODUCTS