7
Measurement Specifications
Color: RM3542A only
Resistance measurement accuracy
Conditions of guaranteed accuracy
Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year
30 minutes or more
Warm-up time
Longer than the default value for the Integration Time Setting Function
(No regulation for settings in ms if the default value is set to PLC)
23°C ±5°C (73°F ±9°F), 80% RH or less
Integration time
Temperature and humidity range for
guaranteed accuracy
Temperature fluctuation after self-calibration must be within ±2°C (±3.6°F). Add Temperature Coefficient ±(1/10 of measurement accuracy)/°C for the
following ranges: 0°C to 18°C (32°F to 64°F) and 28°C to 40°C (82°F to 104°F).
LOW POWER: OFF
Measurement Accuracy: ± (%rdg. + % f.s.)
Maximum
Measurement
Current*2
Open-Circuit
Voltage
Range
Resolution
Display Value*1
FAST
MED
SLOW
100 mΩ
1000 mΩ 1200.000 mΩ
120.0000 mΩ
100 nΩ
1 µΩ
0.015 + 0.008
0.012 + 0.003
0.012 + 0.003
0.010 + 0.003 0.008 + 0.002 0.008 + 0.001
0.009 + 0.003 0.007 + 0.002 0.007 + 0.001
0.009 + 0.003 0.007 + 0.002 0.007 + 0.001
0.008 + 0.003 0.006 + 0.002 0.006 + 0.001
0.009 + 0.003 0.007 + 0.002 0.007 + 0.001
0.009 + 0.003 0.007 + 0.002 0.007 + 0.001
0.010 + 0.003
0.010 + 0.003
0.010 + 0.003 0.008 + 0.002 0.008 + 0.001
0.010 + 0.003 0.008 + 0.002 0.008 + 0.001
0.030 + 0.004
0.015 + 0.003
0.012 + 0.002
0.012 + 0.002
0.015 + 0.002
0.012 + 0.001
0.012 + 0.001
100 mA
100 mA
33.3 mA
10 mA
10 mA
3.33 mA
1 mA
3 Ω
10 Ω
100 Ω
300 Ω
1000 Ω
10 kΩ
30 kΩ
100 kΩ
300 kΩ
1000 kΩ
3 MΩ
10 MΩ
30 MΩ
100 MΩ
3.60000 Ω
12.00000 Ω
120.0000 Ω
360.000 Ω
10 µΩ
10 µΩ
100 µΩ
1 mΩ
1 mΩ
10 mΩ
100 mΩ
100 mΩ
1 Ω
1 Ω
10 Ω
10 Ω
100 Ω
100 Ω
1200.000 Ω
12.00000 kΩ
36.0000 kΩ
120.0000 kΩ
360.000 kΩ
1200.000 kΩ
3.60000 MΩ
12.00000 MΩ
36.0000 MΩ
120.0000 MΩ
1 mA
20 V max
*3, *4, *5
333 µA
100 µA
33.3 µA
10 µA
3.33 µA
1 µA
0.007 + 0.002 0.007 + 0.001
0.007 + 0.002 0.007 + 0.001
0.030 + 0.010
0.100 + 0.020
333 nA
100 nA
LOW POWER: ON
Maximum
Display Value*1
1000 mΩ 1200.000 mΩ
Measurement Accuracy: ± (%rdg. + % f.s.)
Measurement
Current*2
10 mA
3.33 mA
1 mA
Open-Circuit
Voltage
Range
Resolution
FAST
MED
SLOW
1 µΩ
10 µΩ
10 µΩ
100 µΩ
1 mΩ
0.010 + 0.008 0.008 + 0.003 0.008 + 0.002
0.010 + 0.008 0.008 + 0.003 0.008 + 0.002
0.010 + 0.008 0.008 + 0.003 0.008 + 0.002
0.010 + 0.003 0.008 + 0.002 0.008 + 0.001
0.010 + 0.003 0.008 + 0.002 0.008 + 0.001
0.020 + 0.003 0.008 + 0.002 0.008 + 0.001
3 Ω
10 Ω
3.60000 Ω
12.00000 Ω
120.0000 Ω
360.000 Ω
1200.000 Ω
10 V max
*3, *5
100 Ω
300 Ω
1000 Ω
1 mA
333 µA
100 µA
1 mΩ
*1Negative values can be up to 10% of positive full scale.
*2Measurement current accuracy is ±5%.
*3Voltage when not measuring is 20 mV or less, with current mode set at PULSE and Contact Improver Setting set at OFF/PULSE (with a voltmeter having 10 MΩ).
*4VOLTAGE LIMIT ON: 10 V max
*5With the sum of resistances of the probes, sample, and contacts less than (open-circuit voltage) / (measurement current).
Example. 100 mA measurement current can be used when the sum of resistances of the probes, sample, and contacts is no more than 20 Ω.
▲
300
±3
▲
Dimensions (Units: mm)
Recommended Model for
Each Type of Measurement
Related Products
Resistance Meter for the Ultra-Low Shunt Era RM3543
Small
Large
Chip Size
Power Inductor
Ferrite Beads
● Inspection of 0.1mΩ at a high accuracy of 0.16%, and a high resolution
of 0.01 μΩ. Shunt resistor load inspection with superior accuracy and
resolution.
● Excellent repetitive measurement accuracy
● Intuitive user interface and superb noise immunity ideal for use with
automated equipment
RM3542A
RM3542
Common Mode Filter
Chip Inductor
General-Purpose Resistor
Shunt Resistor
RM3543
RM3543
RM3543-01 (With GP-IB)