200 Watts
xppower.com
PBM200 Series
•
•
•
•
•
•
•
Worldwide Medical Approvals
Single, Dual, Triple and Quad Outputs
Small Size - Only 7” x 4” x 1.5”
Power Fail Detect & Remote On/Off
Universal AC Input with Active PFC
High Reliability - 350k Hours MTBF
100 Watt Convection-Cooled Rating
Specification
Input
General
Input Voltage
Input Frequency
Input Current
• 90-264 VAC
• 47-63 Hz
Efficiency
Isolation
• Up to 80%
• 4000 VAC Input to Output
1500 VAC Input to Ground
500 VAC Output to Ground
• 3.2 A rms at 115 VAC,
1.6 A rms at 230 VAC
Switching Frequency
Power Density
• 100 kHz 12 kHz
• 4.76 W/In3
Inrush Current
Power Factor
• 20 A at 115 VAC, 40 A at 230 VAC
cold start 25 ºC
• 0.98 typical
Power Fail Detect
• AC OK = TTL logic high
AC not OK = TTL logic low
Earth Leakage Current • 90 µA max at 115 VAC, 60 Hz
150 µA max at 230 VAC, 50 Hz
MTBF
• 350 kHrs per MIL-HDBK-217F at +25 ˚C
Input Protection
• Both line and neutral fused
Environmental
Output
Operating Temperature • 0 ˚C to +70 ˚C, derate linearly from 100%
power at +50 ˚C to 50% power at +70 ˚C
Output Voltage
Output Voltage Trim
Initial Set Accuracy
Minimum Load
Start Up Delay
Start Up Rise Time
Hold Up Time
• See table
Cooling
• ‘C’ versions have internal fan,
‘B’ versions are derated with convection
cooling or require 25 CFM (see note 4)
• Not user-adjustable
•
2%
• See table & note 5
• 2 s max
Operating Humidity
Storage Temperature
Operating Altitude
Shock
• Up to 95% RH, non-condensing
• -40 ˚C to +85 ˚C
• 18 ms typical
• 2500 m
• 20 ms minimum at full load & 110 VAC
• Half sine pulse. acceleration 50 g, 11 ms
duration, 3 shocks in each direction,
unpackaged, non-operating
Drift
•
•
0.2%
Line Regulation
Load Regulation
Transient Response
0.5% max
• See note 7
Vibration
• Sinusoidal 10-55 Hz, amplitude 0.15 mm,
30 min/axis, unpackaged, non-operating
• 4% max deviation, recovery to within
1% in 500 µs for a 25% load change
EMC & Safety
Ripple & Noise
• 2% max pk-pk, 20 MHz BW with a 22 µF
electrolytic & a 0.47 µF tantalum capacitor
Emissions
• EN55011 & FCC Class B conducted
& radiated
Overvoltage Protection • 112-140% Vnom output 1 only, recycle
input to reset
Harmonic Currents
Voltage Flicker
ESD Immunity
Radiated Immunity
EFT/Burst
• EN61000-3-2
Overtemperature
Protection
• Shutdown at +85 ºC (singles) or +105 ºC
(multi outputs) with auto recovery,
measured internally
• EN61000-3-3
• EN61000-4-2, level 3 Perf Criteria A
• EN61000-4-3, level 3 Perf Criteria A
• EN61000-4-4, level 3 Perf Criteria A
• EN61000-4-5, level 3 Perf Criteria A
• EN61000-4-6 3 V, Perf Criteria A
Overload Protection
• 140-200% typical, auto recovery
Short Circuit Protection • Trip and restart (Hiccup mode)
Surge
Temperature
Coefficient
• 0.04%/˚C
Conducted Immunity
Dips & Interruptions
• EN61000-4-11, 30% 10 ms,
60% 100 ms, >95% 5000 ms,
Perf Criteria A, B, B
Remote Sense
• Single output models only.
Compensates for up to 0.5 V drop
Remote On/Off
• On = Logic low or open circuit
Off = Logic high
Safety Approvals
• EN60601-1, UL60601-1,
CSA22.2 No. 601-1 per cUL, CE Mark