MU425S
425W Single Output
Medical Power Supply
Features
.
.
.
.
.
.
.
3.3” x 6.2” x 1.62” Package
Up to 425W of Air-Cooled Power, 300W Convection
Universal Input 90-264Vac Input Range
5V at 2A Standby Output
12V Fan Output
Inhibit, Power Fail, DC OK Signals, Remote Sense
Approved to CSA/EN/IEC/UL60601-1, 3rd Edition
2 MOPP Isolation
.
Meets EN61000-4-2, EN61000-4-3 and EN61000-4-6
Requirements for Home Healthcare Applications
.
.
.
Efficiency 88% typical
3-Year Warranty
RoHS Compliant
Description
The MU425 family is designed to power the latest professional healthcare devices. All models meet 2 MOPP input-to-output
isolation. Models are available with main output voltages of 12V, 18V, 24V and 48V.
Model Selection
Model
Output Current
Ripple & Noise³
(mV pk-pk)
120mV
Total
Regulation
±3%
Efficiency4
OVP
(Main Output)
Number
Volts
12V
w/200LFM air¹
Convection²
Threshold
14.0 ± 1.1V
5.5V – 8.0V
N/A
21.0 ± 2.0V
5.5V – 8.0V
N/A
28.0 ± 2.5V
5.5V – 8.0V
N/A
58.0 ± 2.0V
5.5V – 8.0V
N/A
88%
88%
90%
90%
32.2A
2.0A
1.0A
22.0A
MU425S12E
MU425S18E
MU425S24E
MU425S48E
5Vsb
2.0A
0.5A
100mV
360mV
±5%
12V Fan6
±10%5
±3%
18V
5Vsb
21.45A
2.0A
1.0A
14.6A
2.0A
0.5A
180mV
100mV
360mV
240mV
100mV
360mV
480mV
100mV
360mV
±5%
12V Fan6
±10%5
±3%
24V
5Vsb
16.8A
2.0A
1.0A
11.9A
2.0A
0.5A
±5%
12V Fan6
±10%5
±3%
48V
5Vsb
8.4A
2.0A
1.0A
5.9A
2.0A
0.5A
±5%
12V Fan6
±10%5
Notes:
1. Total power with 200 lfm of forced air cooling is 425W (408W for 18V, 385W for 12V) including 12V/1A for Fan output and 5V/2A standby.
2. Total convection power is 300W (280W for 12V model). Total power includes 5Vsb and Fan outputs.
3. Measured at 25°C ambient with noise probe directly at end of 6” twisted pair terminated with 0.1μF ceramic and 10μF low ESR capacitors. Values will be higher at
ambient temperatures below 0°C.
4. Efficiency values listed are typical and are measured at 115Vac input, full load output current, at an ambient temperature of 25°C.
5. Fan output regulation is with 1A min load on main output.
6. Fan Output: If the load on the output is other than a fan, a short circuit condition on the output can only be remedied by removing both the cause of the short circuit and
the load. This will allow the output to resume normal operation.
General Specifications
Main output: <1 sec. max @115Vac, rise time 30mS max.
5Vsb turn-on time is 500mS max., rise time 50mS max.
Output Voltage rise is monotonic.
Turn On Time
(Main Output)
100-240Vac, ±10%, 47-63Hz, 1∅
AC Input
120–300Vdc (external fuse required for DC input)
Main Output: >20ms for 300W @ 120Vac/60 Hz, >16ms
for 367W (90% of 408W) @ 120v/60Hz.
5Vsb Output: >500mS
115Vac: 5.2A, 230Vac: 2.5A
Input Current
Hold-up Time
Overtemperature
Protection
264Vac, cold start: will not exceed 40Arms
within ½ cycle. I²T = 25A²/Sec maximum
Sensing transformer temperature, 135°C (55°C ambient
temperature at full load), auto recovery.
Inrush Current
Input Fuses
Efficiency
Overload
Protection
Switching
Frequency
F1, F2: 6.3A, 250Vac
See chart above.
130 to 170% of rating, Hiccup Mode, auto-recovery.
75kHz, typical
www.slpower.com
MU425S REV 3.1
14-APR-16
1