ME60
60W Single Output External Power
Medical Series
Features
.
Meets UL/EN/IEC60601-1-2, 4th edition for
EMC*
.
.
Approved to EN/IEC/UL60601-1, 3rd edition
with isolation levels which satisfy the 2 MOPP
requirements
Meets DoE Efficiency Level VI Requirements
.
.
No load input power
Average Efficiency
.
.
.
.
Up to 60W of AC-DC Power
Universal Input 90-264Vac Input Range
Desktop Style Package
Meets EN55011/CISPR11, FCC Part 15.109
Class B Conducted & Radiated Emissions, with
6db margin
.
.
.
E-cap life of >7 years
3 Year Warranty
IP22 Rated Enclosure
Description
A high performance AC to DC external power supply family designed for medical applications. The ME60A Medical Series low
power external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition and EN/IEC/UL60601-1-11:2010
for Home Healthcare (non-hospital use) applications with and isolation levels which satisfy the 2 MOPP requirements and
designed to UL/EN/IEC60601-1-2, 4th edition for EMC*. The ME60A Series models will operate at universal input range of 90 to
264Vac over the wide temperature range of -20°C to +50°C, delivering full rated output power up to +40°C and applicable
output power derating at 50°C. These models are available in desktop versions, include an IP22 rating per IEC60529 for the
enclosure, and the output cable can be terminated at a variety of output connectors.
*Consult Factory for Table 9 compliance information.
Model Selection
Model
Number
Output
Current
7.00A
Output
Power
35W
Ripple &
Noise¹
75mV pk-pk
Line
Regulation
±1%
Load
Regulation
±5%
Output
Connector
6 pin Molex Type2
Input
Configuration
Volts
5.0V
ME60A0551F01
ME60A1203F01
ME60A1503F01
ME60A1803F01
ME60A2403F01
ME60A4803F01
ME60A0551N01
ME60A1203N01
ME60A1503N01
ME60A1803N01
ME60A2403N01
ME60A4803N01
ME60A0551Q01
ME60A1203Q01
ME60A1503Q01
ME60A1803Q01
ME60A2403Q01
ME60A4803Q01
12.0V
15.0V
18.0V
24.0V
48.0V
5.0V
5.00A
4.00A
3.30A
2.70A
1.35A
7.00A
5.00A
4.00A
3.30A
2.70A
1.35A
7.00A
5.00A
4.00A
3.30A
2.70A
1.35A
60W
60W
60W
60W
60W
35W
60W
60W
60W
60W
60W
35W
60W
60W
60W
60W
60W
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
480mV pk-pk
75mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
480mV pk-pk
75mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
480mV pk-pk
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
Class I Desktop,
IEC60320 C14
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
6 pin Molex Type3
12.0V
15.0V
18.0V
24.0V
48.0V
5.0V
Class II Desktop,
IEC60320 C8
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
6 pin Molex Type3
12.0V
15.0V
18.0V
24.0V
48.0V
Class II Desktop,
IEC60320 C18
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Notes:
1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. For 5V
models, values listed are typical, 100mV pk-pk maximum.
2. Molex p/n 39-01-2060 or equivalent. See outline drawing for pinout information.
3. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (TE60B1203F01).
4. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
www.slpower.com
ME60 Rev 2.4
31-JAN-16
1