15 Watt Global Performance Medical Switcher

GSM15 数据手册

 浏览型号GSM15的Datasheet PDF文件第2页 
GSM15 Medical  
15 Watt Global Performance Medical Switcher  
GLOBAL PERFORMANCE SWITCHERS  
FEATURES:  
D C P o w e r S u p p l i e s  
• Industrys smallest 15 W medically approved switcher  
• Compact size (3.00” x 2.10” x 0.92”)  
• Wide-range ac input: 90-264 Vac  
• Less than 75 μA leakage current @ 120 Vac  
• Approved to UL2601-1, EN60601-1  
• EMI to FCC, CISPR 11 Class B  
• Overvoltage protection standard  
• RoHS compliant models available (G suffix)  
marked to LVD  
SPECIFICATIONS  
Temperature Coefficient  
0.03% / °C typical.  
Ac Input  
90-264 Vac, 47-63 Hz single phase. Class I or class II grounding.  
EMI/EM Compliance  
Input Current  
All models include built-in EMI filtering to meet the following EMC requirements of  
IEC601-1-2.  
Maximum input current at 90 Vac, 60 Hz with full rated output load not to  
exceed 0.6 A.  
Performance Requirement  
Conducted Emissions  
EMC Standard  
Typical Margin  
Input Protection  
Internal ac fuse provided on all units. Designed to blow only if a catastrophic  
failure occurs in the unit -- Fuse does not blow on unsustained overload or  
short circuit.  
EN55011, Class B; FCC Class B 2 dB Class II Gnd  
6 dB Class I Gnd  
Surge Discharge  
EN61000-4-2, Level 3  
EN61000-4-3, Level 3  
EN61000-4-4, Level 3  
EN61000-4-5, Level 3  
2 kV  
2 V  
Inrush Current  
RF Field Susceptibility  
Fast Transients/Bursts  
Surge Susceptibility  
Inrush is limited by internal thermistors. The inrush at 240 Vac, averaged over  
the first ac half-cycle under cold start conditions will not exceed 37 A.  
500 V  
500 V  
25%  
5%  
Efficiency  
69-85% depending on model.  
Conducted RF Susceptibility EN61000-4-6  
Overload Protection  
Fully protected against short circuit and output overload. Short circuit protec-  
tion is cycling type power limit. Factory set to begin power limiting at 23 W.  
Voltage Sags & Surges  
EN61000-4-11  
Medical Safety Approvals  
All models are Certified to be in compliance with the applicable requirements of  
UL2601-1, IEC60601-1, CSA-C22.2 No. 601-1, EN60601-1.  
Overvoltage Protection  
Built in OVP on all models. Approximately 120-140% of output voltage.  
Leakage Current  
The maximum leakage current for GSM15 series will be as follows;  
Output Noise  
0.5% rms, 1% Pk-Pk, 20 MHz Bandwidth, differential mode. Measured with  
noise probe directly across output terminals of the power supply.  
132Vac/60Hz UL2601-1 test method  
Transient Response  
GND  
Connection  
Normal  
Single Fault  
Main Output - 500 µs max. response time for return to within 0.5% of final  
value for a 50% load step change, ∆i/ ∆t< 0.2 A/μs. Maximum voltage devia-  
tion is 3.5%.  
Class I  
75 μA  
39 μA  
105 μA  
54 μA  
Class II  
264Vac/50Hz IEC60601-1 test method  
GND  
Hold-up Time  
10 ms minimum from loss of ac input voltage at full load, nominal line (120  
Vac).  
Connection  
Normal  
Single Fault  
Class I  
Class II  
128 μA  
66 μA  
180 μA  
94 μA  
All specifications are typical at nominal input, full load at 25OC unless otherwise stated  

与GSM15相关器件

型号 品牌 获取价格 描述 数据表
GSM15-12 SLPOWER

获取价格

15 Watt Global Performance Medical Switcher
GSM15-15 SLPOWER

获取价格

15 Watt Global Performance Medical Switcher
GSM15-15G SLPOWER

获取价格

AC-DC Regulated Power Supply Module, 15W
GSM15-24 SLPOWER

获取价格

15 Watt Global Performance Medical Switcher
GSM15-28 SLPOWER

获取价格

15 Watt Global Performance Medical Switcher
GSM15-28G SLPOWER

获取价格

AC-DC Regulated Power Supply Module, 15W
GSM15-5 SLPOWER

获取价格

AC-DC Regulated Power Supply Module, 15W
GSM15-5G SLPOWER

获取价格

AC-DC Regulated Power Supply Module, 15W
GSM15DRAH ETC

获取价格

CONN EDGE DUAL FMALE 30POS 0.156
GSM15DRAI ETC

获取价格

CONN EDGE DUAL FMALE 30POS 0.156