GLD150 GOLD PERFORMANCE MEDICAL SWITCHERS
150 Watt Single Output
Features:
• Compact 4.5” x 7” x 1.7” size
• Power factor corrected to IEC 1000-3-2 Class A
• Less than 300 µA leakage
• EMI compliance to CISPR11, FCC Class B
• Power fail and remote sense standard
• Approved to UL2601, IEC601-1 and CSA22.2
No. 601.1-M90 and 60601: 1990
• 2 year warranty
•
marked to LVD
SPECIFICATIONS
Ac Input
EMI/EMC Compliance
85-264 Vac, 47-63 Hz single phase.
Input Current
2.8 A line current maximum, at 90 Vac, 60 Hz with full rated load,
power factor .99 typical, .96 minimum. Input current harmonic
content meets the requirements of IEC1000-3-2.
Output Power
150 W with convection cooling, 180 W with fan cooling.
Efficiency
Minimum 80% at full rated load with 230 Vac Input. Approximately 3%
less at 115 Vac.
Hold-Up Time
All models include built-in EMI filtering to meet the EMC require-
ments of IEC601-1-2. Unless otherwise stated, all tests are
done at full load and 115 and 230 Vac input.
Conducted Emissions
Static Discharge
RF Field Susceptibility
Fast Transients / Bursts
Surge Susceptibility
Conducted RF Susceptibility EN61000-4-6
Voltage Sags & Surges
EN55011, Class B; FCC Class B
EN61000-4-2, Level 3
EN61000-4-3, Level 3
EN61000-4-4, Level 3
EN61000-4-5, Level 3
EN61000-4-115%
Outputs will remain within regulation limits for 25 ms minimum from
loss of ac input at full load, 10 ms before Power Fail indication.
Dc Ouput
Total regulation is the maximum deviation from the nominal voltage for
all steady state loading conditions. Peak ratings are for 60 s
maximum duration, 10% duty cycle.
Overload Protection
Fully protected against short circuit and output overload. Short circuit
protection is cycling type power limit.
Minimum Load
No minimum load required to maintain output specifications.
Output Noise
0.5% rms, 1% pk-pk, 20 MHz Bandwidth, differential mode. Mea-
sured with noise probe directly across output terminals of
the power supply.
Inrush Current
Inrush 240 Vac is less than 37 A, averaged over the first ac half-
cycle under cold start conditions. Limiting provided by internal
thermistors.
Fan Output
An additional 12 Vdc, 250 mA output suitable for powering a dc fan
is included in all models. The fan output is both current limited and
thermally protected.
Thermal Shutdown
Provided as a standard feature. Designed to protect unit from
prolonged over temperature.
Power Fail
TTL / CMOS compatible output goes low (<0.5 V) 8 ms before
output voltage drops more than 4% below nominal voltage
upon loss of ac power.
Transient Response
Power Good
Main Output - 500 µs typical response time for return to within 0.5%
of final value for a 50% load step change, ∆i/∆t< 0.2
A µs. Maximum voltage deviation is 3%.
Remote Sense
Provided as a standard feature on all models.
Overvoltage Protection
Built in on all models.
Voltage Adjustment
Output Voltage is adjustable +/- 5% with user adjustable potentiometer.
Temperature Coefficient
TTL / CMOS compatible output goes high more than 100 ms after
V1 reaches regulation and should assure that sufficient energy is
stored in the input section to provide normal power fail/shutdown.
Medical Approvals
All models are Certified to be in compliance with the applicable
requirements of UL2601, CSA 22.2 No. 601.1-M90, IEC 601-1
(1988), EN 60601-1: 1990.
Leakage Current
70 µA, 132 Vac @ 60 Hz normal conditions. Single fault conditions,
130 µA, 254 Vac @ 50 Hz.
0.03% / °C typical on all outputs.
Design Verification Documents
Turn On Time
Less than 1 second.
Overshoot
The “Gold” series has undergone rigorous review and design
analysis. The following product documentation is available upon
request;
1. MTBF study
2. DVT Data
3. EMC / Susceptibility test results
Less than 2% overshoot at turn-on under all conditions, less than 1%
overshoot at turn-off under all conditions.
Internal ac fuses provided on both line and neutal on all models.
Inhibit
Inhibit signal is pulled to the V1 output common to reduce average
output voltage to less than 5% of nominal.