5秒后页面跳转
GLC110-215G PDF预览

GLC110-215G

更新时间: 2024-11-30 19:08:39
品牌 Logo 应用领域
SLPOWER 输出元件
页数 文件大小 规格书
2页 266K
描述
AC-DC Regulated Power Supply Module, 110W

GLC110-215G 技术参数

生命周期:Obsolete包装说明:,
Reach Compliance Code:unknown风险等级:5.84
模拟集成电路 - 其他类型:AC-DC REGULATED POWER SUPPLY MODULE认证:CSA, EN, UL
效率(主输出):85%高度:33.02 mm
最大输入电压:264 V最小输入电压:90 V
长度:177.8 mm最高工作温度:50 °C
最低工作温度:最大输出电流:7.3 A
标称输出电压:15 V保护:OUTPUT OVER CURRENT; OUTPUT OVER VOLTAGE
纹波电压(主输出):0.05303 Vrms子类别:Power Supply Modules
最大总功率输出:110 W宽度:107.95 mm
Base Number Matches:1

GLC110-215G 数据手册

 浏览型号GLC110-215G的Datasheet PDF文件第2页 
GLC110 Commercial/GLM110 Medical  
110 Watt Global Performance Switchers  
GLOBAL PERFORMANCE SWITCHERS  
FEATURES:  
• Cost-effective power source  
• Single- and dual-output units  
• Low height (1.30”)  
• 2-year warranty  
• Power fail signal standard  
• Commercial Approved to UL60950-1, EN60950-1,  
CSA22.2 No. 950-95 3rd Ed.  
• Medical Approved to UL2601-1, EN60601-1,  
CSA-C22.2 No. 601.1  
• Complies with EN61000-3-2 Class A  
• RoHS Compliant Model Available (G suffix)  
SPECIFICATIONS  
Ac Input  
90-264 Vac, 47-63 Hz single phase.  
Transient Response  
Input Current  
Main Output: 500 μs typical response time for return to within 0.5% of final value  
for a 50% load step change, ∆i/∆t<0.2 A/μs. Maximum voltage deviation is 3.5%.  
Startup/shutdown overshoot less than 3%. Turn-On Time less than 1 second at 120  
Vac, 25.  
Maximum input current at 90 Vac, 60 Hz with full rated output load not to  
exceed 2.9 A.  
Hold-up Time  
20 ms minimum from loss of ac input at full load, nominal line (120 Vac).  
Overvoltage Protection  
Built in on main output.  
Output Power  
Normal continuous output power is 75 W for unrestricted natural convection,  
or 110 W with 26 cfm air flow.  
Efficiency  
72-85% depending on model.  
Output Regulation  
Input Protection  
Load regulation on dual output models is measured by 40% load change  
from 60% rated load and input voltage change from minimum to maximum  
ratings.  
Internal ac fuse provided on all units. Designed to blow only if a catastrophic  
failure occurs in the unit.  
Inrush Current  
Output #1 requires 1A minimum load for proper regulation of other outputs.  
Initial set tolerance is measured with all outputs at 60% of full rated load.  
Load regulation for single-output models measured by changing load from  
5% to 50% load or 50% to full load in either direction.  
Inrush limited by internal thermistors. Inrush at 240 Vac, averaged over the first ac  
half-cycle under cold start conditions will not exceed 37 A.  
Temperature Coefficient  
0.03%/°C typical on all outputs.  
Voltage Adjustment  
Output #1 adjustment is 5%. Note: output #1 must not be more than 1%  
below nominal to achieve full output regulation on output #2. High voltage  
settings may degrade the reliability of the unit due to excessive power  
EMI/EMC Compliance  
All models include built-in EMI filtering to meet the following emissions require-  
ments:  
dissipation in some outputs.  
EMI SPECIFICATIONS  
COMPLIANCE LEVEL  
Overload Protection  
Fully protected against short circuit and output overload. Short circuit protec-  
tion is cycling type power limit. Factory set to begin power limiting at 120 W.  
Conducted Emissions GLC110  
Conducted Emissions GLM110  
Static Discharge  
RF Field Susceptibility  
Fast Transients/Bursts  
Surge Susceptibility  
Line Frequency Harmonics  
EN55022 Class B; FCC Class B  
EN55011 Class B: FCC Class B  
EN61000-4-2, 6 kV contact, 8 kV air  
EN61000-4-3, 3 V/meter  
EN61000-4-4, 2 kV, 5 kHz  
EN61000-4-5, 1 kV diff., 2 kV com.  
EN61000-3-2 Class A  
Output Noise  
0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential mode. Measured with  
noise probe directly across output terminals of the power supply.  
POWER FAIL  
Power fail signal is TTL or CMOS compatible (output goes low <0.5 V) 5 ms  
before output voltage drops more than 4% below nominal voltage upon  
loss of AC power. The signal is factory set to trip on 84 to 94 Vac brown-out  
depending upon incoming line impedance and distortion. Other settings are  
available to the user through adjustment of built in potentiometer (consult  
factory for assistance). Output will stay low for 20 ms minimum.  
Commercial Safety  
All GLC models are approved to UL60950-1, CSA22.2 No. 950-95 3rd Ed, and  
EN60950-1.  
Medical Safety  
All GLM models are approved to UL2601-1, CSA-C22.2 No. 601.1, EN60601-1.  
All specifications are typical at nominal input, full load at 25OC unless otherwise stated  

与GLC110-215G相关器件

型号 品牌 获取价格 描述 数据表
GLC110-24 SLPOWER

获取价格

110 Watt Global Performance Switchers
GLC110-24G SLPOWER

获取价格

AC-DC Regulated Power Supply Module, 110W
GLC110-524 SLPOWER

获取价格

110 Watt Global Performance Switchers
GLC110-524G SLPOWER

获取价格

AC-DC Regulated Power Supply Module, 110W
GLC120 GSG

获取价格

N Channel Power MOSFET FOR LOGIC DRIVER
GLC12010 ETC

获取价格

SURFCOIL SMT INDUCTORS
GLC12110 ETC

获取价格

SURFCOIL SMT INDUCTORS
GLC130 GSG

获取价格

N Channel Power MOSFET FOR LOGIC DRIVER
GLC140 GSG

获取价格

N Channel Power MOSFET for Logic Drive
GLC15010 ETC

获取价格

SURFCOIL SMT INDUCTORS