DTMPU130 Series
130 Watt, Single Output
Medical Grade Approvals
AC/DC Desktop Power Supply
Rev C
FEATURES
● 100% Burn-in
● Class I Insulation
● Optional On/Off Switch
● IEC-320-C14 Input Inlet
● Output Voltage Protection
● Active Power Factor Correction
● Energy Star 2.0, Efficiency Level V
● Optional Output Connectors Available
● Ultra-Low Leakage Current (under 0.1mA)
● Wide Input Voltage 90 to 260VAC, 47 to 63Hz
● Input Surge Current, Over Voltage, and Over Load Protection
DESCRIPTION
The DTMPU130 Series of AC/DC medical approved power supplies provide 130 Watts of continuous output power. All supplies are Energy
Star 2.0 Level V and UL94V-1 min compliant. This series has single output models with an input voltage range of 90~260VAC. These
supplies also have an IEC-320-C14 AC input for worldwide applications. This series is best suited for use in hospital instrumentation and
many other applications. All models meet FCC Part-18 class B, CISPR-11 EN55011 class B emission limits and are designed to comply
with UL/c-UL (UL 60601-1), TUV/T-mark (EN 60601-1) and new CE requirements. All models are 100% burn-in tested.
SPECIFICATIONS: DTMPU130 Series
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
INPUT (Vin)
TEST CONDITIONS
Min
Nom
Max
Unit
Operating Voltage Range
Input Frequency
90
47
260
63
1.32
0.66
30
50
0.1
2
VAC
Hz
A
A
A
A
mA
s
Input Current (Low Line)
Input Current (High Line)
Inrush Current (Low Line)
Inrush Current (High Line)
Safety Ground Leakage Current
Start-Up Time
Io = Full Load, Vin = 115VAC
Io = Full Load, Vin = 230VAC
Io = Full Load, 25°C, Cool Start, Vin = 115VAC
Io = Full Load, 25°C, Cool Start, Vin = 230VAC
Io = Full Load, Vin = 240VAC / 60Hz
Io = Full Load, Vin = 100VAC
OUTPUT (Vo)
Output Voltage
Load Regulation
See Table
VDC
%
Vin = 230VAC
5
Line Regulation
Io = Full Load
0.5
1
%
Output Power Range
Output Current Range
Ripple & Noise (peak to peak)
Transient Response Time
Hold-Up Time
Vin = 90 to 264VAC
0
130
W
A
%
ms
ms
%/°C
See Table
Full Load, Vin = 90VAC
1
4
Io = Full Load to Half Load, Vin = 100VAC
Io = Full Load, Vin = 110VAC
All Outputs
16
-0.04
Temperature Coefficient
PROTECTION
+0.04
Over Voltage Protection
Over Current Protection
GENERAL
112
110
132
150
%
%
Efficiency
Io = Full Load, Vin = 230VAC
Primary to Secondary
88
40
%
Dielectric Withstanding Voltage
For Primary to Secondary
Dielectric Withstanding Voltage
For Primary to Ground
Isolation Resistance
Power Factor Correction
ENVIRONMENTAL
Operating Temperature
Storage Temperature
Relative Humidity
5656
2800
VDC
Primary to Ground
VDC
Io=Full load, Vin =230VAC
Io = Full Load, Vin = 230VAC
50
0.95
MΩ
1
Derate linearly from 100% Load at 40°C to 50% load at 70°C
Operating Temperature at 25°C, Calculated per MIL-HDBK-217F
0
-40
5
+70
+85
95
°C
°C
%
MTBF
100,000 hours
PHYSICAL
Weight
Dimensions (L x W x H)
SAFETY
Approx. 27.44 ~ 28.22oz (778 ~ 800g)
7.44 x 3.52 x 1.79 inches
EMI Requirements for CISPR-11
EMI Requirements for FCC PART-18
Vin = 220VAC
Vin = 110VAC
B
B
Class
Class
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797