DTM65-C Series
40W to 65W Medical AC-DC External Power Supplies
CEC & EISA(1) Compliant Models
Wide Range AC Input
Medical Certifications
Features and Benefits
Feature
CEC and EISA Compliant
Wide Range AC Input
> 85% Efficient
Benefit
Easier system compliance
Global Operation
Consumes less energy
Specifications
ITEMS
Input Voltage range
Inrush Current
Input Current
Power Factor
90 - 264VAC (47 - 63Hz)
<60A at 230VAC input, 25°C ambient cold start
A
A
2A at 90VAC, Full Load
None
Leakage Current (264VAC 63Hz) mA
<300uA
Hold Up Time (Typ)
Temperature Coefficient
Voltage Accuracy
Adjustment Range
Minimum Load
ms
16ms at 115VAC input
±0.05%/°C
%
±1%
None
None
A
Total Regulation
5V to 15V Models:±5%, >15V models ±3%
Ripple & Noise
%
-
V
%
1% or 100mV whichever is the greater
Continuous - hiccup mode
110 - 130% of nominal (Cycle input power to reset)
Short Circuit Protection
Overvoltage Protection
Efficiency
>85%, average load for 15V and greater models, >75% <15V models
Operating Temperature
Storage Temperature
Humidity (non condensing)
Cooling
Withstand Voltage
Vibration (non operating)
Shock
Safety Agency Approvals
California Energy Commission (CEC)
EISA
0 to +50°C
-10 to +70°C
10 - 90% RH
Convection
-
-
Input to Ground 1.5kVAC, Input to Output 4kVAC (Output is connected to Ground)
23.52m/s2 (10 - 55Hz: constant sweep 1 min X, Y, Z for 1 hour)
< 196.1 m/s2 (20G)
UL60601-1, EN60601-1, IEC60601-1
See model selector
See model selector
Offload Power Consumption
Conducted & Radiated EMI
Immunity
< 0.5W
EN55011, EN60601-1-2 Class B
EN55024
Weight (Typ)
Size (WxLxH)
Cable Length
g
in
mm
460g
2.72 x 5.2 x 1.57
1050mm
AC Input Connector
Output Connector
Warranty
IEC 320-C14 (Accepts IEC 320-C13)
Kycon KPP-4P or equivalent, Pins 1 & 2: -Vout, Pins 3 & 4: +Vout
One Year
(1) EISA - Energy Independence and Security Act of 2007
35
3055 Del Sol Blvd • San Diego, CA 92154 • 1-800-LAMBDA-4