PSMAD100 SERIES
Rev C
Wall Industries, Inc.
100 Watt, Medical AC/DC Power Supplies
SPECIFICATIONS: PSMAD100 SERIES
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Typ
Max
Unit
ENVIRONMENTAL SPECIFICATIONS
Operating Ambient Temperature
Storage Temperature Range
Operating Altitude
Convection cooled: 100W (with derating)
-25
-40
+80
+85
°C
°C
5000 meters
Thermal Shock
IEC68-2-27
IEC68-2-6
Vibration
Relative Humidity
Non-condensing
5
90
% RH
BELLCORE TR-NWT-000332 Case 1: 50% Stress, Ta=40°C
MIL-HDBK-217F, Ta=25°C, full load (G/B, controlled environment)
1,919,000 hours
276,000 hours
MTBF
PHYSICAL SPECIFICATIONS
Open Frame Models (Suffix “-O”)
U-Chassis Models (Suffix “-U”)
Enclosed Case Models (Suffix “-C”)
Open Frame Models (Suffix “-O”)
U-Chassis Models (Suffix “-U”)
Enclosed Case Models (Suffix “-C”)
5.50oz (156g)
6.84oz (194g)
7.41oz (210g)
Weight
3.00 x 2.00 x 1.16 in (76.2 x 50.8 x 29.5 mm)
3.60 X 2.44 X 1.54 in (91.4 x 62.0 x 39.2 mm)
3.60 X 2.44 X 1.54 in (91.4 x 62.0 x 39.2 mm)
Dimensions (L x W x H)
SAFETY & EMC
Safety Approvals
ANSI/AAMI ES60601-1, IEC60601-1, and EN60601-1 3rd edition
Conducted
Class B
Class A
EMI
EN55011, EN 55022 and FCC Part 18
EN61000-3-2 Full load
Radiated
Harmonic Currents
Voltage Flicker
ESD
Class A and Class D
EN61000-3-3
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
Air 8kV and Contact 6kV
Perf. Criteria A
Perf. Criteria A
Perf. Criteria A
Perf. Criteria A
Perf. Criteria A
Perf. Criteria A
Perf. Criteria A
Perf. Criteria A
Perf. Criteria B
Perf. Criteria A
Perf. Criteria B
Perf. Criteria A
Perf. Criteria B
Radiated Immunity
Fast Transient
Surge
10 V/m
2kV
2kV
Conducted Immunity
10 Vrms
30%
500ms
100ms
10ms
60%
>95%
>95%
30%
EN60601-1-2
EN60601-1-2
230VAC 50Hz
100VAC 50Hz
5000ms
500ms
100ms
10ms
Dip and Interruptions
60%
>95%
>95%
5000ms
MODEL NUMBER SETUP
PSMAD
Series Name
100
Output Power
-
12
Ouptut Voltage
S
Output Quantity
-
O
Package Type
B
Protection Type
100: 100 Watts
12: 12 VDC
15: 15 VDC
24: 24 VDC
28: 28 VDC
36: 36 VDC
48: 48 VDC
S: Single Output
O: Open Frame
U: U-Chassis
None: Class I
B: Class II
C: Enclosed Case
D: DIN Rail (1)
(1) DIN Rail option is only available for enclosed case models.
Wall Industries, Inc. • 5 Watson Brook Road, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: sales@wallindustries.com
05/22/2013
Page 3 of 8