GLC65 Commercial/GLM65 Medical
65 Watt Multiple Output Global Performance Switchers
FEATURES:
• 65 Watt Wide Range Input
• High Efficiency (Up to 90%)
• Universal input 90-264 Vac
• 2-year warranty
• Also available in single outputs
• Conducted EMI exceeds FCC Class B and CISPR 22
Class B (Commercial models) and CISPR 11 Class B
(Medical models)
• Exempt from line harmonics standard EN61000-3-2
• Commercial Approved to UL1950, CSA-C22.2 No. 950,
EN60950
• Medical Approved to UL2601-1, CSA 22.2 No. 601.1,
and EN60601-1
marked to LVD
•
SPECIFICATIONS:
Ac Input
90-264 Vac, 47-63 Hz single phase
Minimum Load
Output Power
No minimum load required on any output. However,
regulation limits may be exceeded if extreme conditions
are applied. Contact factory for assistance.
Total continuous output power is 60 W, 75 W peak for 60 s
10% duty cycle. Total continuous rating with 150 LFM of
air is 70 W.
Temperature Coefficient: 0.03%°C typical on all outputs.
Input Current
Units are exempt from input current harmonic require-
ments of EN61000-3-2 when output power is less than
62W. Maximum input current at minimum input voltage
and output overload will be less than 2.0 A.
Voltage Adjustment
Units provided with potentiometer capable of a minimum
adjustment range of +/- 5% from nominal setting.
EMI / EMC Compliance:
Inrush Current
All models include built-in EMI filtering to meet the EMC
requirements of IEC601-1-2.
Inrush is limited by internal thermistor. The inrush at 240
Vac, averaged over the first ac half-cycle under cold start
conditions will not exceed 37 A.
Performance
EMC Standard
Conducted Emissions GLC65 EN55022, Class B; FCC Class B
Conducted Emissions GLM65 EN55011, Class B; FCC Class B
Static Discharge
RF Field Susceptibility
Fast Transients / Bursts
Surge Susceptibility
Input Protection
EN61000-4-2, Level 3
EN61000-4-3, Level 3
EN61000-4-4, Level 3
EN61000-4-5, Level 3
Internal AC fuse provided on all units. Designed to blow
only if a catastrophic failure occurs in the unit -- Fuse
does not blow on short circuit or unsustained overload.
Conducted RF Susceptibility EN61000-4-6, 3V, 80%
HoldupTime
Voltage Sags & Surges
EN61000-4-11
Output voltage stays within regulation for 20 ms from loss
of ac input at 65 W load, from 120 Vac input.
Line Frequency Harmonics
EN61000-3-2 Class A (>62Wout)
Exempt (<62Wout)
GLC65 Commercial - Safety
Efficiency
Condor DC Power Supplies, Inc. declares under our sole
responsibility that all models are in conformity with the
applicable requirements of EN60950 following the
provisions of the Low Voltage Directive 73/23/EEC. All
models are approved to UL1950 (with no D3 deviations),
CSA22.2 No. 950-95 3rd edition, Level 3, IEC950,
EN60950. CB certificate available.
80% typical.
Overload Protection
Fully protected against short circuit and output overload.
Short circuit protection is cycling type power limit and will
automatically recover after removal of fault.
Overvoltage Protection
Built in with firing point set. OVP firing reduces output to
less than 50% of nominal voltage in less than 50 ms.
Commercial Leakage Current
Less then 500 µA @ 120 Vac. Less then 1.2 mA @ 240 Vac.
GLM65 Medical - Safety
Output Noise
Approvals: All models are certified to be in compliance
with the applicable requirements of UL2601-1, CSA 22.2
No. 601.1, IEC 601-1 (1988), EN 60601-1: 1990.
0.5% rms, 1% pk-pk, 20 MHz Bandwidth, differential
mode. Measured with scope probe directly across output
terminals of the power supply with load terminated with
1µF capacitor.
Medical Leakage Current
40 µA under normal conditions (132 Vac @ 60 Hz).
Maximum under single fault conditions (264 Vac @ 60
Hz) is 170 µA.
Transient Response
Main Output - 500 µs typical response time for return to
within 0.5% of final value for a 50% load step within the
regulation limits of minimum and maximum load, ∆i/∆t<
0.2 A/µs. Maximum voltage deviation is 3.5%.