CSS150 Series
www.uk.tdk-lambda.com/css150
3 x 5” 150W AC-DC
Medical Power Supplies
,QGXVWULDO
7HVW
%URDGFDVW
0HGLFDO
&RPP
Featuresꢀ
ꢀ Benefitsꢀ
• No Load Power <0.5W
• Energy Saving
• Low Profile, Industry Standard Footprint • Fits 1U Applications
• Dual input fuses
• Simplifies System Design, Reduces Cost
Specification
Model
CSS150
Input Voltage range
Inrush Current
V
A
A
-
µA
ms
-
%
A
%
%
-
90 - 264VAC (47 - 400Hz)(1) or 120 - 180VDC
<60A maximum at 264VAC input, 25°C ambient cold start
Input Current (115/230VAC)
Power Factor Correction
Leakage Current
Hold Up Time (Typ)
Temperature Coefficient
Voltage Accuracy
Minimum Load
Total Regulation
Ripple & Noise
Overcurrent Protection
Overvoltage Protection
Remote On/Off
1.7 / 0.85
Meets EN61000-3-2, >0.9
<250µA 264VAC 63Hz
16ms at 115VAC input
±0.05%/°C
±1%
None
±3%
2% peak to peak
110 - 150%
V
-
110 - 150% of nominal (Cycle input power to reset)
Unit on: Floating or low <1.2V, Unit off: Apply 5V with respect to 0V
Efficiency
No load power
Operating Temperature
Storage Temperature
Humidity (non condensing)
Cooling
%
W
°C
°C
-
89% typical
<0.5W
0 to +70°C derate linearly to 50% load from 50 to 70°C
-20°C to +85°C
20 - 95% RH
Convection or forced air (2.5m/s)
-
Withstand Voltage
Isolation Resistance
Vibration (non operating)
Shock
-
Ω
-
Input to Ground 1.5kVAC, 3kVAC (2 x MOOPs 3rd Edition), Output to Ground 500VDC for 1 min.
>100M Ω at 25°C & 70%RH, Output to Ground 500VDC
19.6m/s2 (10~55Hz:2G Constant, X,Y,Z 60min each.)
-
< 196.1m/s2 (20G)
Safety Certifications and Markings
Conducted & Radiated EMI
Immunity
-
-
-
IEC/EN/CSA60601-1, ANSI/AAMI ES60601-1, IEC/EN/UL/CSA62368-1, 60950-1, CE Mark & UKCA Mark
EN55011-B, FCC Class B
EN60601-1-2
MTBF
-
180,000 (MIL-217F-HDBK)
Weight (Typ)
g
340g
Size (WxLxH)
Warranty
mm
yrs
127 x 76.2 x 33 (including underside components)
2
(1) Derate linearly to 90W convection rating from 90 to 100VAC input
1