Pressure sensors
Differential pressure transmitter (voltage output)
MiniCell Series
16)
Line pressure pLine
Line pressure pline is defined as the maximum common mode pressure working on both pressure ports simultaneously.
Line pressure above ambient pressure may lead to an additional total error of the sensor.
17)
18)
19)
Line overpressure pov line
1000 pressure cycles within the pressure range 0 up to pov line working on both pressure ports simultaneously will not
affect the performance of the pressure sensor.
Line burst pressure pburst line
Line bust pressure pburst line is the maximum of permissible pressure working on both pressure ports simultaneously
without causing leackage of the sensor. Measurement performance of the sensor may be affected.
Total measuring error ET
Total measuring error ET includes offset error, span error, nonlinearity, pressure hysteresis, temperature hysteresis, and
signal noise. It describes the deviation of the signal to the nominal output characteristic.
TPS PRS T PD
2021-10-21
Please read Cautions and warnings and
Page 8 of 11