30-60 Watts
AFM Series
xppower.com
•
•
•
•
•
•
•
Energy Efficiency Level V
CEC2008 & EISA2007 Compliant
IT & Medical Approvals
China Compulsory Certification (CCC) Qualified
Optional Class II Versions
Optional White Case Versions
3 Year Warranty
Specification
Input
General
Efficiency
Energy Efficiency
• See table
• Level V
Input Voltage
• 80-264 VAC, derate output power 25%
<90 VAC
Isolation
• 4000 VAC Input to Output, 2 x MOPP
1500 VAC Input to Ground, 1 x MOPP*
Output 0 V is electrically connected to
Input Ground*. *Not C8 version.
Input Frequency
Input Current
• 47-63 Hz
• AFM30: 0.8 A rms at 115VAC,
AFM45: 1.2 A rms at 115VAC,
AFM60: 1.5 A rms at 115 VAC
• AFM30: 50 A max at 240VAC,
AFM45: 50 A max at 240VAC,
AFM60: 120 A max at 240VAC
Switching Frequency
Power Density
• 30/45 W: 58 kHz typ., 60 W: 52 kHz typ.
Inrush Current
Power Factor
3
3
• AFM30: 2.8 W/In , AFM45: 4.2 W/In ,
3
AFM60: 3.76 W/In
MTBF
• >250 kHrs to MIL-HDBK-217F at 25 °C,
GB, full load
• EN61000-3-2, class A
Earth Leakage Current • 150 µA max at 264 VAC/60 Hz
No Load Input Power • AFM30/45: <0.3 W, AFM60: <0.5 W
Environmental
Operating Temperature • 0 °C to +60 °C, derate from 100% power
at +40 °C to 60% power at +60 °C
• 10-95% RH, non-condensing
• -40 °C to +80 °C
Output
Operating Humidity
Storage Temperature
Operating Altitude
Shock
Output Voltage
• See table
• Not user-adjustable
Output Voltage Trim
Initial Set Accuracy
Minimum Load
• 2000 m
•
2%
• 30 g, 10 ms on 3 axes
• No minimum load required
• 3 s max at 115 VAC
2
Vibration
• 5-100 Hz, 2.31 m/s , 20 mins, 3 axes
Start Up Delay
Hold Up Time
• 8 ms minimum at 115 VAC
EMC & Safety
Emissions
Line Regulation
Load Regulation
Transient Response
•
1.0% maximum
• EN55011/22/FCC Class B conducted,
EN55011/22/FCC Class B radiated
• EN61000-3-2, Class A
• EN61000-3-3
• EN61000-4-2, 6 kV contact and
8 kV air discharge
• 2.5% maximum
• 4% max. deviation, recovery to within
1% in 500 µs for a 25% load change
Harmonic Currents
Voltage Flicker
ESD Immunity
(1)
Ripple & Noise
• 1% max pk-pk
Overvoltage Protection • 110-170% Vnom, recycle input to reset
Overload Protection
• AFM45/60: 110-170%,
AFM30: 110-220%, auto recovery
Radiated Immunity
EFT/Burst
• EN61000-4-3, 3 V/m Perf Criteria A
• EN61000-4-4, level 3 Perf Criteria A
• EN61000-4-5, installation class 3
Perf Criteria A
• EN61000-4-6, 3 Vrms Perf Criteria A
• EN61000-4-8, 3 A/m, Perf Criteria A
(7)
Short Circuit Protection • Trip and restart (hiccup mode)
Temperature
Coefficient
Surge
• 0.04%/°C
Conducted Immunity
Magnetic Field
Dips & Interruptions
• EN61000-4-11, 30% for 10 ms,
60% 100 ms, 100% 5000 ms,
Perf Criteria A, B, B
EN60601-1-2, 30% 500 ms, 60% 100 ms,
100% 10 ms, 100% 5000 ms,
(8)
Perf A, A, A, B
Safety Approvals
• EN60950-1, UL60950-1, IEC60950-1
EN60601-1 (3rd Ed.), UL60601-1 (2nd Ed.),
IEC60601-1 (3rd Ed.) China Compulsory
Certification (CCC) Qualified.