Easily monitor leak current fluctuations
In combination with a HIOKI MEMORY HiCORDER this unit can be used for unattended monitoring for leak current condition.
Example for a monitoring system using the 3283 and 8807 MEMORY HiCORDER
The 3283 can supply a REC signal (recording signal, 1V DC/f.s.)
9199
or a MON signal (waveform signal, 1V AC/f.s.). The 9094
OUTPUT CABLE (option) or conversion 9199 BNC TO
BANANA ADAPTER (option) can be used for connection to
the 8807. For monitoring of instantaneous leak current
waveforms, set the output to MON, and select a suitable trigger
level in the memory mode of the 8807. This will cause
recording to be performed only when leak current is detected.
The trigger output terminal (open collector output) can be used
as an alarm output. For long-term monitoring, the 8807 should
be powered from the 9418-10 AC ADAPTER and the 3283
from the 9445-02 AC ADAPTER .
BNC TO BANANA ADAPTER
8807
MEMORY HiCORDER
9198
CONNECTION
CORD (for Voltage input)
9094 OUTPUT CORD
(cord length approx.1.5m)
Setting example for 3283 and 8807
When leak current exceeds 20 mA, the trigger is activated and recording is carried
out, extending also to data before triggering occurred. The 3283 is set to the 100
mA range, and the output is set to MON.
8807 waveform printout example
Trigger point
MON
RMS
mA
0
3
6
9
3283 indication
8807 trigger screen
Pre-trigger
✔✔3283 Specification (accuracy at 23˚C±3˚C 80% rh)
✔✔Measurable conductor diameter: ø40mm max. ✔✔Maximum
usable circuit voltage: 300Vrms (insulated conductor) ✔✔Functions:
recording (MAX. MIN. AVE), data hold, auto power-off ✔✔Display
update rate: digital indication FAST approx. 4 times/second,
NORMAL approx. 2 times/second, SLOW 1 time/ 3 seconds; bar
graph indication approx. 4 times/second ✔✔Filter: ON 180Hz±30Hz(-
3dB) ✔✔External output (A function only): REC (recording signal,
1VDC/f.s. ±3% rdg.±10mV) circuit time constant 200ms or less /
MON (waveform signal, 1VAC/f.s. ±3%rdg.±10mV, frequency
bandwidth 5Hz to 15kHz), selectable (2V/f.s. for 200A range) ✔
Effect of conductor position: within ±0.1% in any direction from the
center of sensor ✔✔Effect of external magnetic field: 400A/m
corresponds to 5mA, max. 7.5 mA ✔✔Crest factor: 2.5max. (1.5 for
200A range) ✔✔Safety standard: EN61010-1, EN61010-2, EN60529
Function
Range
Accuracy
Notes
10.00mA
±1.0%rdg.±5dgt.
(45Hz to 66Hz)
Maximum
allowable current
100.0mA
1.000 A
10.00 A
±2.0%rdg.±5dgt.
(40Hz to 45Hz, 66Hz to 2kHz)
20Arms cont.
AC current
AC A
±1.5%rdg.±5dgt.
(45 to 66Hz)
Maximum
allowable current
200.0 A
±2.0%rdg.±5dgt.
(40 to 45Hz, 66 to 2kHz)
200Arms cont.
Accuracy range
±0.3%rdg.±1dgt.
±1.0%rdg.±1dgt.
Frequency 100.0Hz
30.0 to 99.9Hz
95 to 1000Hz
Hz
1000Hz
AC current measurement accuracy range is 10% to 100% (5% to 100% in 200 A
range). Accuracy is given for the filter OFF condition.
IP40 ✔✔EMC: EN61326-1 ✔✔Power supply: 9445 AC ADAPTER or
!
WARNING
6F22 manganese battery✔1 (continuous operation max. 50 hours) ✔
Dimensions and mass: Approx. 62W✔225H✔39D mm; 400 g
Inspect the unit and check that it is operating correctly before use. When carrying out
measurement on live lines, wear proper protective gear, insulating rubber gloves, insulating
rubber boots and safety helmet, and use extreme caution to avoid electric shock accidents.
3283 CLAMP ON LEAK HiTESTER
(All include 9399 CARRYING CASE, HAND STRAP)
! DANGER
Option
1. In order to prevent short-circuits and injury, use the clamp product on electrical
circuits with a voltage less than the maximum operation circuit voltage.
2. In order to prevent short-circuits and injury when the clamp core tip is open, do
not use bare conductors.
9445-02 AC ADAPTER (for the America, Japan)
9445-03 AC ADAPTER (for the EU type)
*
9094 OUTPUT CORD (Jack - banana plug)
*No CE marking
DISTRIBUTED BY
HEAD OFFICE :
Shanghai Representative Office :
1704 Shanghai Times Square Office
93 Huaihai Zhong Road
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@hioki.co.jp
Shanghai, 200021, P.R.China
TEL +86-21-6391-0090, 0092
FAX +86-21-6391-0360
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@hiokiusa.com
E-mail: info@hioki.cn
All information correct as of Dec. 22, 2004. All specifications are subject to change without notice.
3283E3-4ZE-02K Printed in Japan