Datasheet
Corrected system performance with calibration kit
Tektronix TCAL500 35mm
SMA-type electrical calibration
kits
(TCAL500-35F, TCAL500-35MF, TCAL500-35M)
Frequency
Directivity Source
Load
match
(dB)
Reflection Transmission tracking
tracking (±dB)
(±dB)
(dB)
match
(dB)
300 kHz to < 500 kHz
500 kHz to < 2MHz
2 MHz to < 10 MHz
10 MHz to < 200 MHz
200 MHz to < 1.5 GHz
1.5 GHz < 3 GHz
-47.5
-50.7
-50.7
-50.7
-46.6
-46.6
-39.7
-36.6
-39.6
-39.6
-39.6
-34.8
-34.8
-25.3
-47
0.062
0.062
0.062
0.062
0.08
0.066
0.042
0.042
0.042
0.073
0.073
0.156
-47
-47
-47
-37.1
-37.1
-36
0.08
3 GHz to 6 GHz
0.166
18 °C to 28 °C, within 1 °C of calibration temperature and at the same ambient humidity conditions in which calibration was
performed.
Tektronix TCAL500 N-type
electrical calibration kits
(TCAL500-NF, TCAL500-NMF, TCAL500-NM)
Frequency
Directivity Source
Load
match
(dB)
Reflection Transmission tracking
tracking (±dB)
(±dB)
(dB)
match
(dB)
300 kHz to < 500 kHz
500 kHz to < 2MHz
2 MHz to < 10 MHz
10 MHz to < 200 MHz
200 MHz to < 1.5 GHz
1.5 GHz < 3 GHz
-45.8
-45.8
-45.8
-45.8
-41.5
-41.5
-35.7
-33.8
-36.3
-39.1
-39.1
-34.8
-34.8
-25.3
-47
0.062
0.062
0.062
0.062
0.08
0.066
0.042
0.042
0.042
0.073
0.073
0.156
-47
-47
-47
-37.1
-37.1
-35
0.08
3 GHz to 6 GHz
0.166
18 °C to 28 °C, within 1 °C of calibration temperature and at the same ambient humidity conditions in which calibration was
performed.
Spinner Type-N mechanical
calibration kit BN533861
User Correction ON.
Two 60 cm Tektronix cables 012-1765-00 or 012-1768-00
Frequency
Directivity Source
Load
Load
Reflection Transmission tracking:
(dB)
match
(dB)
match:
match with tracking
Insertable devices (dB)
Insertable M-M or F-F (dB)
devices
(dB)
thru
(dB)
300 kHz to < 1 MHz
1 MHz to < 10 MHz
10 MHz to < 100 MHz
100 MHz to < 1 GHz
1 GHz to < 3 GHz
3 GHz to 6 GHz
-38
-37
-37
-36
-36
-36
-34
-34
-34
-34
-34
-34
-37
-37
-37
-37
-37
-36
-35
-35
-35
-35
-35
-35
0.08
0.08
0.08
0.08
0.08
0.09
0.05
0.02
0.01
0.01
0.02
0.02
18 °C to 28 °C, within 1 °C of calibration temperature and at the same ambient humidity conditions in which calibration was
performed.
6 www.tek.com