Figure 2-4 Single Level Internal Planar Probe  
Figure 2-5 Single Level Internal Spiral Wire Probe  
T2  
T2  
Figure 2-6 Twin-Level Internal Planar Probe  
Figure 2-7 Twin-Level Internal Spiral Wire Probe  
Spiral wire probes have the disadvantage of not being as Coaxial probes are more expensive to make, and can have  
rugged as a solid disc probe.  
problems with vibration if they are not constructed robustly.  
The outer cylinder should be perforated at key spots to allow  
fluid to fill and drain the cavity without trapping air bubbles  
inside. The outer cylinder can also be made of a wire mesh.  
2.2.3 SIDE-ENTRY  
PROBES  
Another type is a side-entry probe (Figure 2-8), which  
requires an entry point into the vessel wall, but may have the  
advantage of accessibility in certain cases. These can be  
made of simple metal rod, insulated in almost any plastic if  
required.  
The outer cylinder does not have to be coated in plastic, even  
when used with water-based fluids. When used with oils, the  
inner rod does not require insulation either.  
2.2.5 BARE  
M
ETAL  
PROBES  
2.2.4 COAXIAL  
PROBES  
Bare metal internal probes can be used, for example with  
nonconductive fluids like oils, without difficulty. This applies  
to all probe types described above.  
Another type of internal probe is the coaxial probe (Figure  
2-10); these are most useful with oils or similar fluids having  
a low dielectric constant; the inner rod is connected to the  
signal connection, and together with the outer grounded  
cylinder forms a capacitor whose dielectric is either air or oil.  
Keeping the gap between rod and cylinder to a minimum  
increases the 'gain' of the electrode.  
Bare probes can also be used with aqueous fluids, but in  
these cases a 1,000pF (1nF) ceramic NPO capacitor should  
be inserted between the probe and the QT114 to block DC  
current flows.  
LQ  
4
QT114 R1.04/1106  

与QT114_06相关器件

型号 品牌 描述 获取价格 数据表
QT114-D QUANTUM CHARGE-TRANSFER QLEVEL SENSOR IC

获取价格

QT114-DG QUALCOMM QLEVEL⑩ SENSOR IC

获取价格

QT114-IS QUANTUM CHARGE-TRANSFER QLEVEL SENSOR IC

获取价格

QT114-ISG QUALCOMM QLEVEL⑩ SENSOR IC

获取价格

QT114-S QUANTUM CHARGE-TRANSFER QLEVEL SENSOR IC

获取价格

QT115 QUANTUM QT 11x Series Variations

获取价格