5秒后页面跳转
16445A-001 PDF预览

16445A-001

更新时间: 2024-01-12 22:02:05
品牌 Logo 应用领域
惠普 - HP /
页数 文件大小 规格书
23页 300K
描述
Agilent B1500A Semiconductor Device Analyzer

16445A-001 数据手册

 浏览型号16445A-001的Datasheet PDF文件第1页浏览型号16445A-001的Datasheet PDF文件第2页浏览型号16445A-001的Datasheet PDF文件第4页浏览型号16445A-001的Datasheet PDF文件第5页浏览型号16445A-001的Datasheet PDF文件第6页浏览型号16445A-001的Datasheet PDF文件第7页 
• Multi-frequency capacitance measure-  
ment unit (MFCMU) available  
• High-voltage semiconductor pulse  
generator unit (HV-SPGU) available  
• Waveform generator/fast measurement  
unit (WGFMU) available  
• High-resolution, analog-to-digital  
converter (ADC) available to all installed  
SMUs  
• High-speed ADC present on each  
installed SMU  
• SMU/AUX path switching supported on  
the atto-sense and switch unit (ASU)  
• SMU/MFCMU switching supported  
using SMU CMU unify unit (SCUU) and  
guard switch unit (GSWU)  
• MFCMU automatically identifies  
capacitance measurement accessories  
• WGFMU/SMU path switching supported  
on the remote-sense and switch unit  
(RSU)  
• GPIB port for instrument control  
• Self-test, self-calibration, diagnostics  
The measurement and output accuracy  
are specified at the rear panel connector  
terminals when referenced to the Zero  
Check terminal. The B1530A WGFMU  
measurement and output accuracy are  
specified at the output terminal of the RSU.  
Accuracy is specified under the following  
conditions:  
Basic Features  
• PC-based instrument with touch screen  
interface; optional USB keyboard and  
mouse available  
• EasyEXPERT software with over 200  
categorized application tests supplied  
with the instrument  
• Performs current versus voltage (IV)  
measurements  
• Performs capacitance versus voltage  
(CV), capacitance versus time (Ct)  
and capacitance versus frequency (Cf)  
measurements  
• Performs quasi-static CV (QSCV)  
measurements  
• High voltage pulse forcing up to 40 V  
with ALWG and voltage monitor  
capabilities.  
• Performs fast IV measurement synchro-  
nized with the applied waveforms for  
accurate transient IV or time-domain  
measurements  
• Front-panel Classic Test measurement  
modes supported: single-channel sweep,  
multi-channel sweep, time sampling,  
list sweep, CV sweep and direct control  
(GPIB FLEX)  
• GUI-based control of the Agilent  
B2200A, and B2201A switching matrices  
• Modular mainframe with ten module  
slots and one 4.2 A ground unit  
• Multiple source/monitor unit (SMU)  
types available: medium power  
(MPSMU), high-power (HPSMU), and  
high-resolution (HRSMU)  
1. Temperature: 23 °C 5 °C  
2. Humidity: 20 % to 60 %  
3. After 40 minute warm-up followed by  
self-calibration  
4. Ambient temperature change less than  
1 °C after self-calibration eꢀecution,  
not applicable for MFCMU and WGFMU  
5. Measurement made within one hour  
after self-calibration eꢀecution, not  
applicable for MFCMU and WGFMU  
6. Calibration period: 1 year  
7. SMU integration time setting:  
1 PLC (1 nA to 1A range, voltage range)  
20 PLC (100 pA range)  
50 PLC (1 pA to 10 pA range)  
Averaging of high-speed ADC:  
128 samples per 1 PLC  
8. SMU filter: ON (for SMUs)  
9. SMU measurement terminal connection:  
Kelvin connection  
Specification conditions  
This document lists specifications and  
supplemental information for the B1500A  
and its associated modules. The specifica-  
tions are the standards against which the  
B1500A and its associated modules are  
tested. When the B1500A and any of its  
associated modules are shipped from the  
factory, they meet the specifications. The  
“supplemental” information and “typical”  
entries in the following specifications are  
not warranted, but provide useful informa-  
tion about the functions and performance  
of the instrument.  
10. WGFMU load capacitance: 25 pF or less  
Note: Agilent Technologies is responsible  
for removing, installing, and replacing the  
B1500A modules. Contact your nearest  
Agilent Technologies to install and  
calibrate the B1500A modules  
B1500A specification  
Supported plug-in modules  
The B1500A supports ten slots for plug-in modules.  
Part  
Description  
Slots  
Range of operation  
Measure  
number  
occupied  
resolution  
2 µV, 10 fA  
B1510A High power source/monitor unit (HPSMU)  
2
1
1
1
1
-200 V to 200 V, -1 A to 1 A  
B1511A Medium power source/monitor unit (MPSMU)  
B1517A High resolution source/monitor unit (HRSMU)  
E5288A¹ Atto-sense and switch unit (ASU)  
B1520A Multi frequency capacitance measurement unit (MFCMU)  
B1525A High voltage semiconductor pulse generator unit (HV-SPGU)  
B1530A Waveform generator/fast measurement unit (WGFMU)  
-100 V to 100 V, -100 mA to 100 mA 0.5 µV, 10 fA  
-100 V to 100 V, -100 mA to 100 mA 0.5 µV, 1 fA  
-100 V to 100 V, -100 mA to 100 mA 0.5 µV, 100 aA  
1 kHz to 5 MHz  
40 V (80 Vp-p)  
PG Mode: -3 V to 3 V, -5 V to 5 V  
Fast IV Mode: -3 V to 3 V, -5 V to  
5 V, 0 V to 10 V, -10 V to 0 V  
0.035 fFrms²  
50 μV  
1
0.014 % of the  
range 3  
1. This is connected with the B1517A high resolution SMU.  
2. Dispersion of measurement values when connecting a DUT 10 pF to the measurement terminals under the measurement condition of frequency  
1 MHz, signal level 250 mVac, and measurement time 1 PLC. The display resolution is 0.000001 fF at 1 fF order by 6 digits display.  
3. Display resolution  
2

与16445A-001相关器件

型号 品牌 获取价格 描述 数据表
16445A-002 HP

获取价格

Agilent B1500A Semiconductor Device Analyzer
1644614 PHOENIX

获取价格

HEAVYCON PROTECTIVE COVER
1-644803-0 MACOM

获取价格

MTA-100 SHROUDED HEADER ASSEMBLY
1-644803-1 MACOM

获取价格

MTA-100 SHROUDED HEADER ASSEMBLY
1-644803-2 MACOM

获取价格

MTA-100 SHROUDED HEADER ASSEMBLY
1-644803-3 MACOM

获取价格

MTA-100 SHROUDED HEADER ASSEMBLY
1-644803-4 MACOM

获取价格

MTA-100 SHROUDED HEADER ASSEMBLY
1644818 PHOENIX

获取价格

Connector Accessory,
16-44UD-BD-16 ETC

获取价格

0.16 X 0.44 BD 16--16-44UD-BD-16
16-44UD-MAG-16 ETC

获取价格

0.16 X 0.44 MAG 16--16-44UD-MAG-