30-40 WATT MEDICAL POWER SUPPLIES

TM41-24 数据手册

 浏览型号TM41-24的Datasheet PDF文件第2页 
30-40 WATT MEDICAL POWER SUPPLIES  
TRUMPower  
DESCRIPTION  
(LVD)  
TM41 SERIES  
The TM41 series of compact, open PCB constructed,  
AC-DC switching power supplies are capable of  
delivering 30 to 40 watts of continuous output power.  
They operate at 85 to 264VAC input voltage without  
the need of voltage selection. They are ideally suited  
for use in medical equipments, safety systems and  
monitoring equipments.  
RoHS  
FEATURES  
Low safety ground leakage current  
100% burn-in  
85 to 264VAC wide input range  
Input surge current protection  
Overvoltage protection  
Overcurrent protection  
Safety Standard Approvals  
Open PCB construction  
Compliant with RoHS requirements  
UL 2601-1  
CSA C22.2 No. 601.1  
INPUT SPECIFICATIONS  
85 to 264VAC  
110 to 280VDC  
47 to 63Hz  
1.1Armsfor 115VAC  
Input voltage  
TÜV EN60601-1  
Input frequency:  
Input current:  
0.6Armsfor 230VAC  
90µA max. @ 115VAC, 60Hz  
150µA max. @ 230VAC, 50Hz  
GENERAL SPECIFICATIONS  
Earth leakage current:  
Switching frequency: 42KHz ±5KHz  
Efficiency:  
70% minimum on single output  
model with Vo 12V, 68%  
minimum on the others  
Hold-up time:  
Line regulation:  
Inrush current:  
20 msec minimum at 110VAC  
±0.5% maximum at full load  
OUTPUT SPECIFICATIONS  
Output voltage/current:  
Total output power:  
See rating chart  
12 amps @ 115VAC or 24 amps  
@ 230VAC, at 25ºC cold start  
See rating chart  
Withstand voltage:  
MTBF:  
4000VAC from input to output  
1500VAC from input to ground  
500VAC from output to ground  
Ripple and noise:  
1% peak to peak max.  
Over voltage protectionProvided on output; set at  
112 – 132% of its nominal  
600,000 hours minimum at full load  
at 25ºC ambient, calculated per  
MIL-HDBK- 217F  
output voltage  
Over current protectionThe output protected to short  
circuit conditions  
EMC Performance (EN60601-1-2: 2001)  
Temperature coefficientAll outputs ±0.04%/ºC  
EN55011:  
Class B conducted, Class B radiated  
Transient response:  
Maximum excursion of 4% or  
better on all models,  
FCC:  
Class B conducted, Class B radiated  
Class B conducted, Class B radiated  
Harmonic distortion, Class A and D  
Line flicker  
VCCI:  
recovering to 1% of final value  
within 500µs after a 25% step  
load change  
EN61000-3-2:  
EN61000-3-3:  
EN61000-4-2:  
EN61000-4-3:  
EN61000-4-4:  
EN61000-4-5:  
EN61000-4-6:  
EN61000-4-8:  
ESD, ±8KV air and ±6KV contact  
Radiated immunity, 3V/m for 80-2500MHz  
Fast transient/burst, ±2KV  
ENVIRONMENTAL SPECIFICATIONS  
Operating temperature:  
Storage temperature:  
Relative humidity:  
Derating:  
0ºC to +70ºC  
Surge, ±1KV diff, ±2KV com.  
Conducted immunity, 3Vrms  
-40ºC to +85ºC  
Magnetic field immunity, 3A/m  
5% to 95% non-condensing  
Derate from 100% at +50ºC  
linearly to 50% at +70ºC  
EN61000-4-11: Voltage dips, 30% reduction for 500ms,  
60% reduction for 100ms and >95%  
reduction for 10ms  
Tumbler Technologies + TRUMPower  
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA  
Phone: 408-988-6616 Fax: 408-988-6622 email: sales@trumpower.com Website: www.TRUMPower.com  

与TM41-24相关器件

型号 品牌 描述 获取价格 数据表
TM41-28 TRUMPOWER 30-40 WATT MEDICAL POWER SUPPLIES

获取价格

TM41-5.1 TRUMPOWER 30-40 WATT MEDICAL POWER SUPPLIES

获取价格

TM4161EM4-15L ROCHESTER Video DRAM Module

获取价格

TM4161EM4-15L TI IC,VRAM,FAST PAGE,64KX4,MOS,SIP,31PIN,PLASTIC

获取价格

TM4161EM4-20L ROCHESTER Video DRAM Module

获取价格

TM4161EM4-20L TI IC,VRAM,FAST PAGE,64KX4,MOS,SIP,31PIN,PLASTIC

获取价格