5秒后页面跳转
QT114-ISG PDF预览

QT114-ISG

更新时间: 2024-02-04 19:51:50
品牌 Logo 应用领域
高通 - QUALCOMM 传感器
页数 文件大小 规格书
14页 247K
描述
QLEVEL⑩ SENSOR IC

QT114-ISG 技术参数

是否Rohs认证: 不符合生命周期:Obsolete
包装说明:SOP, SOP8,.25Reach Compliance Code:unknown
风险等级:5.92Is Samacsys:N
商用集成电路类型:CONSUMER CIRCUITJESD-30 代码:R-PDSO-G8
端子数量:8最高工作温度:70 °C
最低工作温度:封装主体材料:PLASTIC/EPOXY
封装代码:SOP封装等效代码:SOP8,.25
封装形状:RECTANGULAR封装形式:SMALL OUTLINE
电源:3 V认证状态:Not Qualified
子类别:Other Consumer ICs表面贴装:YES
技术:CMOS温度等级:COMMERCIAL
端子形式:GULL WING端子节距:1.27 mm
端子位置:DUALBase Number Matches:1

QT114-ISG 数据手册

 浏览型号QT114-ISG的Datasheet PDF文件第3页浏览型号QT114-ISG的Datasheet PDF文件第4页浏览型号QT114-ISG的Datasheet PDF文件第5页浏览型号QT114-ISG的Datasheet PDF文件第7页浏览型号QT114-ISG的Datasheet PDF文件第8页浏览型号QT114-ISG的Datasheet PDF文件第9页 
The trip point ideally occurs at the  
centerline of the internal probe or external  
electrode; this can be trimmed with a  
In the case of coaxial probes, the ground  
connection is inherent in the outer cylinder  
and no further ground connection is  
required.  
potentiometer  
if  
necessary  
(see  
Section 3.2). Making the electrode narrow  
and long (horizontally) will help keep the  
trip point localized within a narrow band.  
3 PROCESSING AND  
CIRCUITRY  
2.4 DUAL LEVEL SENSING  
When two trip levels are desired, for  
example for high-low limit sensing, the  
electrode or probe set should have two  
3.1 SLOSH FILTER  
It is desirable to suppress rapid, multiple  
detections of fluid level generated by the  
surface movement of the fluid, for example  
in a moving vehicle. To accomplish this,  
distinct tiers.  
A typical twin external  
electrode is shown in Figure 2-3 (they are  
connected together to the sense line);  
typical internal twin electrodes are shown  
in Figures 2-6, 2-7, 2-9, and 2-11. The  
response of a properly constructed 2-tier  
probe is shown in Figure 2-3.  
the QT114 incorporates  
a
detection  
integration counter that increments with  
each detection until a limit is reached, after  
which point one of the OUT lines is  
activated. If during a detection ‘event’ the  
fluid level falls below the electrode level  
(signal rises above a 'T' point in signal  
counts), the counter decrements back  
towards zero. Over a long interval the up  
and down counts will tend towards either  
zero or the limit, with the result being a  
statistical function of the number of  
detections vs. nondetections. If on average  
there are more detections than nondetections, the counter  
will eventually make its way to the limit value and an OUT line  
will activate.  
Dual level electrodes should have an  
approximately 3:1 surface area ratio or  
more from T2 to T1; that is, the surface  
area at T2 should be at least 3x the  
surface area of the electrode at T1. There  
is no penalty for making T2 excessively  
large. The high ratio is required to  
Figure 2-12 A two-tier spiral wire  
probe with ground rod  
overcome the QT114's decreasing gain with increasing Cx  
load (Figures 4-1, 4-2).  
With internal dual-level probes where T1 and T2 are  
substantially separated, the intervening connection between  
the two levels should be more thickly insulated, for example  
with a thick plastic spacer, and any remaining internal gap  
inside the spacer should be filled with silicone sealant or  
epoxy. This will help to prevent the signal from rising much  
between the two levels, thus preserving a crisp bi-level  
response like that shown in Figure 2-3.  
Once a detection has been established, the counter must find  
its way back to zero before the affected OUT line goes  
inactive, via the same process. Although the counter has a  
nominal reaction time of 15 seconds, in some cases it may  
take several minutes before the outcome is resolved  
depending on the violence of the fluid surface. If the fluid  
surface is stable however, it will only require 15 seconds to  
change the state of an OUT line.  
2.5 GROUNDING CONSIDERATIONS  
In all cases ground reference coupling to the fluid must be  
made. In aqueous fluids, this can simply mean connecting  
the metal vessel to circuit ground, or inserting a bare metal  
element into the bottom of a plastic or glass vessel. The  
degree of galvanic contact is not critical, so scale and  
corrosion on the ground electrode are not of great concern  
especially if the 'connection' to the fluid is substantial enough .  
Both OUT1 and OUT2 have their own independent slosh  
filters. Both are enabled or disabled in unison by strap option,  
pin 4, 'FILT' as follows:  
FILT = Gnd  
FILT = Vcc  
Slosh filter off  
Slosh filter on  
FILT strapping can be changed 'on the fly'.  
If direct electrical contact to the fluid is not possible, a large  
piece of external metal can be bonded to the outside of the  
vessel and grounded. Once this is done, the signal should be  
monitored while the vessel is touched by hand; if the  
grounding is sufficient, the signal will not move or will move  
only slightly.  
3.2 CALIBRATION  
Both the T1 and T2 trip point values are hardwired internally  
as functions of counts of burst length. Sensitivity can be  
altered relative to these trip points by altering electrode size,  
geometry, degree of coupling to the fluid, and the value of  
Cs. Selecting an appropriate value of Cs for a given electrode  
geometry is essential for solid detection stability.  
Very large vessels, even if not grounded, often do not require  
additional provision for grounding since the bottom surface  
area and free-space capacitance of the tank may be  
sufficient for ground return coupling.  
The QT114 employs dual threshold points set at 250 and 150  
counts of acquisition signal. The signal travels in a reverse  
direction: increasing Cx reduces the signal counts; as a  
result, 250 counts of signal corresponds to the most sensitive  
or ‘lower’ setting (T1), and 150 the least sensitive 'upper'  
setting (T2).  
In some cases (windshield washer tanks on cars for  
example) there will exist a water path to a chassis-grounded  
fitting somewhere downstream of the tank, or the water path  
may be labyrinthine enough to provide enough capacitive  
coupling to the grounded chassis even if it does not make  
galvanic contact. In these cases no further provision for fluid  
grounding is required. Simple experimentation will easily  
determine whether the existing amount of parasitic coupling  
to ground is enough to do the job.  
LQ  
6
QT114 R1.04/1106  

与QT114-ISG相关器件

型号 品牌 描述 获取价格 数据表
QT114-S QUANTUM CHARGE-TRANSFER QLEVEL SENSOR IC

获取价格

QT115 QUANTUM QT 11x Series Variations

获取价格

QT115-D ETC IC-QPROX SENSOR

获取价格

QT115H QUANTUM QT 11x Series Variations

获取价格

QT115-S ETC IC-SMD-QPROX SENSOR

获取价格

QT117L QUANTUM Designed to Output raw Data over a Master-mode SPI Port

获取价格