Medical Grade 225 Watt Enclosed Power Supply with PFC
PMMK225E series
GENERAL SPECIFICATIONS
INPUT SPECIFICATIONS
Isolation (Note 2)
I/P-O/P: 5656VDC
Input Voltage Range
Frequency Range
90-264 VAC (100-240VAC Nom)
47-63 Hz
I/P-Ground: 2121VDC
O/P-Ground: 707VDC
≥20MΩ
Power Factor Correction
Inrush Current, typ:
0.95, min. Nom I/P, FL
Insulation Resistance
Efficiency (FL. Nom I/P)
Safety UL/cUL:
30A @ 115VAC Cold Start, FL*
60A @ 230VAC Cold Start, FL*
3.8 Amps max., 90V I/P
1.3 Amps max., 264V I/P
<0.3mA, 240Vin, 50Hz
See Selection Chart
ANSI/AAMI ES60601-1 3rd ed.
CSA C22.2 No. 60601-1 3rd ed.
EN60601-1 3rd ed.
IEC60601-1 3rd ed.
EN60601-1-2
Input Current
UL-EU:
CB:
Leakage Current
CE:
OUTPUT SPECIFICATIONS
ENVIRONMENTAL SPECIFICATIONS
Voltage and Current
See Selection Chart
Line Regulation (LL-HL)
Load Regulation 10%-FL
Preset Accuracy (Note 3)
Ripple/Noise (Notes 1, & 4)
Over Voltage Protection
Over Current Protection
Short Circuit Protection
±1%, max.
Oper. Temperature
Storage Temperature
Relative Humidity
-20 to +70°C (See Derate Curve)
±1%, max.
-25 to +85°C *
10% to +90%, non-cond.*
Storage: 5-95%
±1%, typ (±2%, 5 Vout)
See Selection Chart
130% max. Auto Recovery *
228% max. Auto Recovery *
Auto Recover after fault
condition is removed *
<4S
EMS
Harmonics
Class D, 61000-3-2
61000-3-3
Fluctuations
ESD
61000-4-2
Turn On Delay Time
Rise Time
Hold Up Time
RS
61000-4-3
<20mS, typ (Nom I/P, FL)
20mS, typ (Nom I/P, FL)
<10% of Nominal Output
Compensates for up to 0.5V drop
Open Collector, Logic “1” = ON
“0” = Off. No Termination necessary
to maintain constant “On” operation
EFT
61000-4-4
Surge
61000-4-5
Over-Shoot/Under Shoot
Remote Sense
CS
61000-4-6
MS
DIPS
61000-4-8
Remote Shutdown
61000-4-11
EMI
EN55022 Class B
79,000 Hrs, min.
4G Peak, 50-500Hz, 3 Axes
70 CM
MTBF Mil Std 217F, 25°C
Vibration
PHYSICAL SPECIFICATIONS
Drop Test
Size
3.86”W x 7.01”L x 1.5”H
Enclosed
Construction
Weight
NOTES
1.54 lb, (700g)
1. All measurements should be made directly at the terminals of the
power supply
2. Isolation for up to 1 minute duration.
3. Preset Accuracy measured at nominal load, 120VAC input.
4. O/P Noise measured directly at the pins/terminals at nominal load,
0.1uF ceramic bypass and 47uF electrolytic, pk-pk @ 20MHz
bandwidth.
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
WWW.ASTRODYNETDI.COM