5秒后页面跳转
M2RS-41-M/N PDF预览

M2RS-41-M/N

更新时间: 2024-02-26 20:15:45
品牌 Logo 应用领域
爱模 - MSYSTEM /
页数 文件大小 规格书
4页 106K
描述
Super-mini Signal Conditioners Mini-M Series

M2RS-41-M/N 数据手册

 浏览型号M2RS-41-M/N的Datasheet PDF文件第2页浏览型号M2RS-41-M/N的Datasheet PDF文件第3页浏览型号M2RS-41-M/N的Datasheet PDF文件第4页 
MODEL: M2RS  
Super-mini Signal Conditioners Mini-M Series  
[1] INPUT RTD (2- or 3-wire)  
1: JPt 100 (JIS'89)  
(Usable range: -200 to +500°C, -328 to +932°F; min.span: 50°C, 90°F)  
3: Pt 100 (JIS'89)  
RTD TRANSMITTER  
Functions & Features  
(Usable range: -200 to +650°C, -328 to +1202°F; min.span: 50°C, 90°F)  
4: Pt 100 (JIS'97, IEC)  
• Accepts direct input from an RTD  
• Linearization  
(Usable range: -200 to +650°C, -328 to +1202°F; min.span: 50°C, 90°F)  
5: Pt 50 Ω (JIS'81)  
• Burnout  
• “Active bridge” circuit containing two constant current  
sources allows large leadwire resistances up to 200 Ω  
• Fast response type available  
(Usable range: -200 to +500°C, -328 to +932°F; min.span: 100°C, 180°F)  
6: Ni 508.4 Ω  
(Usable range: -50 to +200°C, -58 to +392°F; min.span: 30°C, 54°F)  
C: Cu (Refer to ‘Cu INPUT’.)  
Typical Applications  
0: Specify  
• Long distance transmission between the RTD and the  
transmitter  
Note: Consult M-System for 2-wire RTD  
• Combination with intrinsic safety barriers  
[2] OUTPUT  
Current  
23 (.91)  
A: 4 – 20 mA DC (Load resistance 750 Ω max.)  
B: 2 – 10 mA DC (Load resistance 1500 Ω max.)  
C: 1 – 5 mA DC (Load resistance 3000 Ω max.)  
D: 0 – 20 mA DC (Load resistance 750 Ω max.)  
E: 0 – 16 mA DC (Load resistance 900 Ω max.)  
F: 0 – 10 mA DC (Load resistance 1500 Ω max.)  
G: 0 – 1 mA DC (Load resistance 15 kΩ max.)  
Z: Specify current (See OUTPUT SPECIFICATIONS)  
Voltage  
Model  
76  
(2.99)  
ZERO  
SPAN  
124  
(4.88)  
mm (inch)  
1: 0 – 10 mV DC (Load resistance 10 kΩ min.)  
2: 0 – 100 mV DC (Load resistance 100 kΩ min.)  
3: 0 – 1 V DC (Load resistance 1000 Ω min.)  
4: 0 – 10 V DC (Load resistance 10 kΩ min.)  
5: 0 – 5 V DC (Load resistance 5000 Ω min.)  
6: 1 – 5 V DC (Load resistance 5000 Ω min.)  
0: Specify voltage (See OUTPUT SPECIFICATIONS)  
MODEL: M2RS–[1][2]–[3][4]  
ORDERING INFORMATION  
• Code number: M2RS-[1][2]-[3][4]  
Specify a code from below for each [1] through [4].  
(e.g. M2RS-4A-M2/BL/CE/Q)  
[3] POWER INPUT  
AC Power  
• Temperature range (e.g. 0 – 500°C)  
• Special output range (For codes Z & 0)  
• Specify the specification for option code /Q  
(e.g. /C01/V01)  
M: 85 – 264 V AC (Operational voltage range 85 – 264 V,  
47 – 66 Hz)  
(Select ‘/N’ for ‘Standards & Approvals’ code.)  
M2: 100 – 240 V AC (Operational voltage range 85 – 264 V,  
47 – 66 Hz)  
For the input code C, specify also the following. If not  
specified, Cu 10Ω @25°C (usable range -200 – +260°C) will  
be used.  
(90 – 264 V for UL)  
DC Power  
• Input type (e.g. Cu 10Ω)  
R: 24 V DC  
• Resistance range (e.g. 9.038 – 12.891 Ω)  
(Operational voltage range 24 V ±10 %, ripple 10 %p-p max.)  
R2: 11 – 27 V DC  
(Operational voltage range 11 – 27 V, ripple 10 %p-p max.)  
(Select ‘/N’ for ‘Standards & Approvals’ code.)  
P: 110 V DC  
(Operational voltage range 85 – 150 V, ripple 10 %p-p max.)  
(110 V ±10 % for UL)  
M2RS SPECIFICATIONS  
ES-5013 Rev.25 Page 1/4  
http://www.m-system.co.jp/  

与M2RS-41-M/N相关器件

型号 品牌 获取价格 描述 数据表
M2RS-41-M/Q MSYSTEM

获取价格

Super-mini Signal Conditioners Mini-M Series
M2RS-41-M/UL MSYSTEM

获取价格

Super-mini Signal Conditioners Mini-M Series
M2RS-41-M2 MSYSTEM

获取价格

Super-mini Signal Conditioners Mini-M Series
M2RS-41-M2/BL MSYSTEM

获取价格

Super-mini Signal Conditioners Mini-M Series
M2RS-41-M2/CE MSYSTEM

获取价格

Super-mini Signal Conditioners Mini-M Series
M2RS-41-M2/K MSYSTEM

获取价格

Super-mini Signal Conditioners Mini-M Series
M2RS-41-M2/N MSYSTEM

获取价格

Super-mini Signal Conditioners Mini-M Series
M2RS-41-M2/Q MSYSTEM

获取价格

Super-mini Signal Conditioners Mini-M Series
M2RS-41-M2/UL MSYSTEM

获取价格

Super-mini Signal Conditioners Mini-M Series
M2RS-41-P MSYSTEM

获取价格

Super-mini Signal Conditioners Mini-M Series