MODEL: JDL
• Thermocouple linearizable range
USABLE RANGE
T/C
°C
°F
(PR)
K (CA)
E (CRC)
0 to 1760
-270 to +1370
-270 to +1000
32 to 3200
-454 to +2498
-454 to +1832
J
(IC)
-210 to +1200
-270 to +400
0 to 1820
-346 to +2192
-454 to +752
32 to 3308
T (CC)
B (RH)
R
S
-50 to +1760
-50 to +1760
-58 to +3200
-58 to +3200
Note: For the temperatures that range below 0°C, the
transmitter may partially not satisfy the described accuracy.
Consult factory.
• RTD linearizable range
USABLE RANGE
RTD
°C
°F
JPt 100 (JIS ’89)
Pt 100 (JIS ’89)
Pt 100 (JIS ’97, IEC)
Pt 50Ω (JIS ’81)
Ni 508.4 Ω
-200 to +500
-200 to +650
-200 to +650
-200 to +500
-50 to +200
-328 to +932
-328 to +1202
-328 to +1202
-328 to +932
-58 to +392
Note:Pt100(JIS’89)isdeviatedfromPt100(JIS’97) only
within the described accuracy.
INSTALLATION
Power input
•AC: Operational voltage range: rating ±10 %,
50/60 ±2 Hz, approx. 3 VA
•DC: Operational voltage range: rating ±10 %,
ripple 10 %p-p max., approx. 3 W (120 mA at 24 V)
Operating temperature: -5 to +55°C (23 to 131°F)
Operating humidity: 30 to 90 %RH (non-condensing)
Mounting: Surface or DIN rail
Weight: 350 g (0.77 lb)
PERFORMANCE in percentage of span
Accuracy: ±0.1 % with segment gain ≤ 1 [±0.1 % × gain]
with segment gain > 1
Temp. coefficient: ±0.015 %/°C (±0.008 %/°F)
Response time: ≤ 0.5 sec. (0 – 90 %)
Line voltage effect: ±0.1 % over voltage range
Insulation resistance: ≥ 100 MΩ with 500 V DC
Dielectric strength: 2000 V AC @1 minute (input to output
to power to ground)
JDL SPECIFICATIONS
ES-1668 Rev.12 Page 3/4
http://www.m-system.co.jp/