MODEL: 10JF
and parameters, square root extraction, zero and span, etc.
(Refer to the users manual of JXCON for the adjustments
configurable with JXCON.)
EQUATION
Equation parameters
X0: output (%)
X
1
to X
to K
to A
2
: input (%)
K0
2: gain ±29.999
INPUT SPECIFICATIONS
A0
2: bias (%) ±299.99 %
■ DC Current: Input resistor incorporated
EQUATION
/1: Temperature compensation for DP flowmeter (ideal gas)
INSTALLATION
K1X1
X0 =
K2X2+A2
Current consumption: Approx. 60 mA with voltage output 1
Approx. 90 mA with current output 1
Operating temperature: -5 to +55°C (23 to 131°F)
Operating humidity: 30 to 90 %RH (non-condensing)
Mounting: Standard Rack 10BXx
where X0 : compensated flow (linear characteristic)
X1 : uncompensated flow (square root extraction available)
X2 : temperature
/2: Pressure compensation for DP flowmeter (ideal gas)
X0 = K1X1 K2X2+A2
Weight: 220 g (0.49 lb)
where X0 : compensated flow (linear characteristic)
X1 : uncompensated flow (square root extraction available)
X2 : pressure
PERFORMANCE in percentage of span
Accuracy: Input accuracy + output accuracy
Input accuracy: ±0.2 % (gain ≤ 1)
/3: Addition/subtraction
X0 = K0 {K1 (X1+A1) + K2 (X2+A2)} + A0
/4: Multiplication
[±0.2 % × gain] with gain > 1
X0 = K0 (K1X1+A1) (K2X2+A2) + A0
Output accuracy: ±0.2 %
/5:Division
Temp. coefficient: ±0.015 %/°C (±0.008 %/°F)
Response time: ≤ 1.2 sec. (0 – 90 %)
0
K (K1X1+A1)
(K2X2+A2)
X0
=
+ A0
Line voltage effect: ±0.1 % over voltage range
Insulation resistance: ≥ 100 MΩ with 500 V DC
Dielectric strength: 500 V AC @ 1 minute
(input to output 1 to output 2 to power)
1500 V AC @ 1 minute (input or output or power to ground)
RELATED PRODUCTS
• Programming Unit (model: PU-2x)
• PC configurator software (model: JXCON)
Downloadable at M-System’s web site.
A dedicated cable is required to connect the module to the
PC. Please refer to the internet software download site or
the users manual for the PC configurator for applicable
cable types.
GENERAL SPECIFICATIONS
Construction: Rack-mounted; terminal access via screw
terminals at the front and via card-edge connector at the
rear; terminal cover provided
Connection
Input: M3.5 screw terminals (torque 0.8 N·m)
Output: Card-edge connector and M3.5 screw terminals
(torque 0.8 N·m)
Power input: Supplied from card-edge connector
Screw terminal: Nickel-plated steel
Housing material: Flame-resistant resin (black)
Isolation: Input to output 1 to output 2 to power (non-
isolated between inputs)
Overrange input: Approx. -25 to +125 %
Overrange output: Approx. -10 to +120 % at 1 – 5 V
Adjustments: Programming Unit (model: PU-2x); equation
10JF SPECIFICATIONS
ES-0880 Rev.11 Page 2/3
http://www.m-system.co.jp/